RPM4/HPMS A70M/A20M-AF Reference Pressure Monitor with High Pressure Mounting System Operation and Maintenance Manual

4. REMOTE OPERATION
Page 81 © 2007 DH Instruments, a Fluke Company
ARANGEn
Purpose Read or set a new AutoRange range to use.
Command
Query
“ARANGEn Range, unit, mode (,RptLabel) (set to user defined range)
“ARANGE Q-RPTLabel (set to Q-RPT default range)
“ARANGEn?”
Classic
Query
“ARANGEn= Range, units, mode (,RptLabel) (set to user defined range)
“ARANGE= Q-RPTLabel (set to Q-RPT default range)
“ARANGEn
Optional Suffix
“n”
A suffix may be used instead of “RptLabel
‘1’ Specify the Hi Q-RPT
‘2’ Specify the Lo Q-RPT
Range:
The AutoRange maximum pressure. Cannot be negative.
Unit:
The unit of measure of the Range argument.
Mode:
The measurement mode of the Range argument:
“A” for absolute
“G” for gauge
Arguments
RptLabel:
(optional)
One of the Q-RPTs that is available.
“IH”: Hi Q-RPT
“IL”: Lo Q-RPT
Remarks The ARANGE command is used to define an RPM4 range by specifying a desired unit of
measure, measurement mode and maximum pressure, and have the RPM4 pick the best
available Q-RPT and make adjustments to optimize operation in that range.
The optional suffix or the “RptLabel” argument can be used to specify which Q-RPT you
wish to use for the AutoRange range instead of allowing the RPM4 to pick the best Q-
RPT. By using this option you can override the internal logic that picks the most suitable Q-
Q-RPT. This Q-RPT specified must be valid for the maximum pressure and measurement
mode specified.
The reply indicates the current range data, including the Q-RPTLabel used for the range.
If “inWa” is specified for the unit, the temperature reference can be given after the unit
text (“inWa4”, “inWa20” or “inWa60” corresponding to inWa at 4 °C, 20 °C or 60 °F). If no
number is given, a default of 20 °C is assumed. There is no indication of the temperature
reference in the reply.
If you just specify the Q-RPTLabel without any other arguments, the Q-RPT specified will
become active in it’s last range and settings. The previously set unit and measurement
mode for the Q-RPT specified will become active.
Example
(enhanced)
Cmd sent:
Query reply:
“ARANGE?” (read the current range in the current units)
“100.00, psi, A, IH”
Example
(classic)
Cmd sent:
Query reply:
“ARANGE” (read the current range in the current units)
“100.00, psi, A, IH”
Example
(enhanced)
Cmd sent:
Query reply:
“ARANGE? 250, inWa4, G (range of 250 inWa @ 4 °C in gauge mode)
“250.00 inWa, G, IH” (Lo Q-RPT is used)
Example
(classic)
Cmd sent:
Query reply:
“ARANGE=500, kPa, G (range of 500 kPa in gauge mode)
“500.00 kPa, G, IL” (Lo Q-RPT is used)
Example
(enhanced)
Cmd sent:
Query reply:
“ARANGE2 50, psi, A (range of 50 psi in absolute mode on
the Lo Q-RPT)
“50.000 psi, A, IL” (No reply if GPIB-488)
Example
(enhanced)
Cmd sent:
Query reply:
“ARANGE IH (Hi Q-RPT’s previous range made active)
“500.000 kPa, A, IH” (No reply if GPIB-488)
Errors ERR# 6:
ERR#10:
ERR# 19:
ERR# 20:
ERR# 29:
ERR# 53:
Range“ maximum pressure exceeds available Q-RPTs or is negative.
RPT specified by the optional suffix is not valid for the current operating
conditions
Cannot Range to ‘0’ with absolute units.
Cannot Range to ‘0’ with gauge units.
Correct type of Q-RPT for the selected “Mode” is not available.
RPT specified is not available.
See Also “RPT”, “RANGE”, “MMODE”, “UNIT”
3.3.4, .3.2.3, 3.3.3, 3.3.2
ATM
Purpose Reads the next measured pressure from RPM4’s on-board barometer (if present).
Query
Classic
“ATM?”
“ATM”
Remarks The atmospheric pressure as measured by the RPM4 on-board barometer is returned in
the current pressure units of the active Q-RPT (always absolute). This measurement is
followed by the units text. Not all RPM4s are equipped with an on-board barometer.
Example
(enhanced)
Query sent:
Query reply:
“ATM?”
“97.12348 kPa a”
Example
(classic)
Sent:
Reply:
“ATM”
“97.12384 kPa a”
Errors ERR# 23: RPM4 is not equipped with a barometer.
See Also 3.5.7.1, 3.2.2