Datasheet

2 Fluke Corporation Fluke 87V IMSK Industrial Multimeter Service Kit
87V IMSK Specifications
87V Industrial Digital Multimeter
Electrical specifications
DC voltage Range 0.1 mV to 1000 V
Accuracy ±(0.05 % + 1)
Maximum resolution 10 µV
AC voltage Range 0.1 mV to 1000 V
Accuracy ±(0.7 % + 2) True-rms
AC bandwidth 20 kHz with low pass filter; 3 dB @ 1 kHz
Maximum resolution 0.1 mV
DC current Range 0.1 μA to 10 A (20 A for 30 seconds maximum)
Amps Accuracy ±(0.2 % + 2)
Maximum resolution 0.01 µA
AC current
Range 0.1 μA to 10 A (20 A for 30 seconds maximum)
Amps Accuracy ±(1.0 % + 2) True-rms
Maximum resolution 0.1 µA
Resistance
Range 0.1 Ω to 50 MΩ
Accuracy ±(0.2 % + 1)
Maximum resolution 0.1 Ω
Capacitance
Range 1 nF to 9999 μF
Accuracy ±(1 % + 2)
Maximum resolution 0.01 nF
Frequency
Range 0.5 Hz to 199.99 kHz
Accuracy ±(0.005 % + 1)
Maximum resolution 0.01 Hz
Duty cycle
Maximum duty cycle 99.9 %
Accuracy ±(0.2 % per kHz + 0.1 %)
Maximum resolution 0.1 %
Temperature measurement –200.0 °C to 1090 °C
–328.0 °F to 1994.0 °F excluding probe
80 BK temperature probe 40.0 °C to 260 °C, ± 2.2 °C or 2 % whichever is greater
40.0 °F to 500 °F, ± 4.0 °F or 2 % whichever is greater
Conductance
Maximum conductance 60.00 nS
Accuracy ±(1.0 % + 10)
Maximum resolution 0.01 nS
Diode
Range 3 V
Resolution 1 mV
Accuracy ±(2 % + 1)
Duty cycle range
Accuracy Within ±(0.2 % per kHz + 0.1 %)
Data storage
Peak transient capture 250 μS
Min/Max/Avg Yes
Reading hold Yes