IR Settings Menu
Table 9 lists the options available in the IR Settings menu.
Auto Off
LCD Time Out Sets the time before the display automatically turns off.
Power Off Sets the time before the Product automatically turns off.
Date <options> Sets the date format and the date.
Time <options> Sets the time format and the time.
Fluke Logo NA Shows or hides the Fluke logo on the display.
Language <options> Sets the language to use on the display.
Decimal
Separator
<options> Sets the decimal separator to comma or decimal point.
Reset to Factory
Defaults
NA
Erases user-set preferences and restores the factory default
settings.
Certificates NA Shows information about the wireless certificates of the Product.
Licenses NA
Shows information about the Open Source Software Licenses used
to develop the Product.
Version NA Shows information about the version of the Product.
Table 9. IR Settings Menu
Option Description
Emissivity
Select an emissivity value from a list of common materials. Use the scroll bar to
see all of the options. See Emissivity Adjustment.
Background
Changes the background temperature to compensate for reflected background
temperature.
Very hot objects or very cold objects can affect the apparent temperature and
measurement accuracy of the target, especially when surface emissivity is low.
Adjust the reflected background temperature to improve the accuracy of the
measurement.
Transmissivity
Changes the values used to calculate the temperature based on the transmission
percentage of the infrared-transparent window (IR window). See Transmissivity.