Datasheet

SSP-KONSTANTER, Series 62 N and Series 64 N
Computer Controlled Laboratory Power Supplies
2 GOSSEN METRAWATT GMBH
Adjustable Functions
Voltage and current setpoints
Voltage and current limit values (soft limits)
Output activation and deactivation
Overvoltage protection trigger value
Overcurrent protection (limiting with or without shutdown)
Delay time for overcurrent shutdown
Power-on status
Reset device configurations
Save device configurations to memory
Recall device configurations individually or sequentially
Select trigger input function
Rounding of measurement value for display purposes
Service request masks (SRQ) *
Activate and deactivate digital displays *
Start self-test at power-up *
*
Only via computer interface
Acquirable Information
Current voltage and current measurement values
Min. and max. voltage and current measurement values
Current output power
Current device settings
Current device status (control mode, excessive temperature, busy *)
Occurred event (power or phase failure, excessive
temperature, overvoltage, overload, program error *)
–Device ID *
*
Only via computer interface
Protective and Additional Functions
Sensor terminals protected against pole reversal
with automatic switching to sensing mode (auto-sensing)
Protection against excessive temperature
Protection against output pole reversal
Locking of front panel operation
Device configuration memory with battery backup
Power and phase failure recognition
In-rush current limiting
Auto-Sensing
Operation in the sensing mode (remote sensing) can be activated
in order to compensate for voltage drops at the power lines.
The sensing mode is automatically activated by connecting the
(–) negative sensing terminal to the negative load level.
Max. compensatable
voltage drop 1 V per line
Locking of Front Panel Operation
The control elements can be secured against unauthorized use by
pressing a key, with a computer command, or by applying a
signal to the trigger input.
Activating and Deactivating the Output
The power output can be activated and deactivated by pressing a
key, with a computer command, or by applying a signal to the
trigger input (no electrical isolation).
Power-On Status
The power supply can be placed into one of the following
conditions after power-up:
Reset = default setting (0 V, 0 A, output inactive etc.)
Recall = last setting (same as prior to last shutdown)
Standby = last setting, but with inactive output
Overcurrent Protection
The device can be configured to respond to current limiting in one
of the following ways:
OCP off = continuous current limiting (UI characteristic curve)
OCP on = deactivate output if duration of current limiting
is greater than delay time
Delay time: setting range from 0.00 to 99.99 s
resolution: 10 ms
Trigger Selection
The device can be configured to respond to the floating trigger
input at the analog interface in one of the following ways:
output = activate / deactivate the power output
local lock = lock the control elements
recall = individual, step-by-step recall of stored
settings
sequence = start / stop the sequence function