Datasheet

GMC-I Messtechnik GmbH 11
SYSKONP500, P800, P1500, P3000 and P4500
Computer Controlled Laboratory Power Supply
1)
At maximum current setting not including processing time for the previous voltage setting command.
2)
Nominal load: Rload = Uset² / Pnom
Output operating characteristics (ppm and percentage specificati-
ons refer to the respective setting and/or measured value)
Reference Conditions
Ambient
temperature 23 °C 2 K
Relative humidity 40 ... 60 %
Warm-up time 30 minutes
Terminals (rear panel)
Mains input
SYSKON P3000
/4500:
connection terminals
(min. 16 A)
(L1 + L2 + L3 + N + PE)
Output
SYSKON P3000
/4500:
Terminal blocks with thread for M8 and M6
screws and 4 mm dia. holes
Analog interface / Double-row plug connector
sensing leads with two 10-pole screw terminals
General
Power supply with 230 V~ nominal line voltage
Power consumption
Line voltage
At nominal load, 100%
At no load
3x230/400 V~ + 10 /
15 %
47 … 63 Hz
3810 VA; 3710 W
100 VA; 45 W
3x230/400 V~ + 10 /
15 %
47 … 63 Hz
5660 VA; 5500 W
110 VA; 55 W
Power supply with 115 V~ nominal line voltage
Power consumption
Line voltage
At nominal load, 50%
At no load
3x115/200 V~ + 10 /
15 %
47 … 63 Hz
2215 VA; 2180 W
73 VA; 48 W
3x115/200 V~ + 10 /
15 %
47 … 63 Hz
3305 VA; 3255 W
92 VA; 60 W
Max. power loss
At a nominal load
3000 W/4500 W (230 V~)
At a nominal load 1500 W/2250 W (115 V~)
710 W
680 W
1100 W
1030 W
Efficiency
At a nominal load
3000 W/4500 W (230 V~)
At a nominal load 1500 W/2250 W (115 V~)
81 %
69 %
82 %
69 %
Switching frequency, PFC / DC/DC Typical 47 kHz / 230 kHz 47 kHz / 230 kHz
Inrush current Max. 50 A
s
50 A
s
Mains fuse (6.3 x 32 mm, UL) 3 x M 15 A / 250 V 3 x M 15 A / 250 V
MTBF (mean time between failures) at 40 °C > 40,000 hours > 30,000 hours
Article Number K363A K364A
Type SYSKON P3000-060-120 SYSKON P4500-060-180