Datasheet

GMC-I Messtechnik GmbH 5
SYSKONP500, P800, P1500, P3000 and P4500
Computer Controlled Laboratory Power Supply
Analog Interface
Functions – Auto-sensing mode
– 2 programmable trigger inputs
– 3 programmable signal outputs
– Voltage control input (0 ... 5 V)
– Current control input (0 ... 5 V)
– Voltage monitor output (0 ... 10 V)
– Current monitor output (0 ... 10 V)
– Master-slave parallel operation
– Master-slave series operation
– Auxiliary power output: 15 V / 60 mA
Computer Interfaces
IEC-625 / IEEE 488 interface (optional)
RS 232 interface
Transmission mode Half-duplex, asynchronous
Transmission speed 1200 to 115,200 baud, adjustable
•USB port
USB port: 4-pin, type B
USB 1.1 compatible with USB 2.0
Connector pin assignments 1: VCC, 2: D-, 3: D+, 4: GND
Transmission speed 9600 to 115,200 baud, adjustable
Power supply
Line voltage
115
/230 V ~ + 10 / – 15%; 47 to 63 Hz
Starting current Max. 50 A
S
Mains fuse
SYSKON P500/P800/P1500
:
1 x M15 A / 250 V (6.3 x 32 mm), UL
SYSKON P3000/4500: 3 x M15 A/250 V
Electrical Safety
Safety class I
Measuring category II for mains input
I for output and interfaces
Pollution degree 2
Earth leakage current < 2.5 mA
RMS
Electrical isolation Test voltage
Output – mains 2.2 kV ~
Output – bus/ground 1.4 kV ~
Mains – bus/ground 2.2 kV –
Bus – ground No electrical isolation
Environmental Conditions
Temperature range Operation: 0 to 40 °C
Storage: –25 to +75 °C
Atmospheric
humidity Operation: 75% rel. humidity,
no condensation allowed
Storage: 65% rel. humidity
Cooling With integrated fan
(temperature controlled)
Inlet vent: Side panel
Outlet vent: Rear panel
Operating noise Noise pressure level at a distance of 30 cm
with fan set to low / high
Front 17 / 28 dBA
Rear 22 / 32 dBA
Left 17 / 28 dBA
Right 20 / 31 dBA
Electromagnetic Compatibility
SYSKON P500/P800/P1500
Generic standard EN 61326-1: October 2006
Interference emission EN 55022: class B
Interference immunity EN 61000-4-2: feature A
EN 61000-4-3: feature B
EN 61000-4-4: feature A
EN 61000-4-5: feature A
EN 61000-4-6: feature A
EN 61000-4-8: feature A
EN 61000-4-11: feature A
SYSKON P3000/4500
Generic standard EN 61326-1: October 2006
Interference emission EN 55022: class A *
Interference immunity EN 61000-4-2: feature B
EN 61000-4-3: feature A
EN 61000-4-4: feature B
EN 61000-4-5: feature B
EN 61000-4-6: feature A
EN 61000-4-8: feature A
EN 61000-4-11: feature B
* Note:
Approved for the deployment in industrial environment. This
device may cause radio interferences in domestic areas.
Applicated Standards
IEC 61010-1:2010, DIN EN 61010-1:2010, VDE 0411-1:2011
EN 61326
Mechanical Data
Protection IP 00 for device and interface connections
IP 20 for housing
Table Excerpt Regarding Significance of IP Codes
Design Benchtop device, suitable for installation to 19" cabinets
HE =
standard height units
IP XY
(1
st
char. X)
Protection against pene-
tration by solid particles
IP XY
(2
nd
char. Y)
Protection against
penetration by water
0 Not protected 0 Not protected
1 50.0 mm dia.
1 Vertical dripping
2 12.5 mm dia.
2 Dripping (15 inclination)
Article No. Designation Dimensions (W x H x D) Weight
K346A SYSKON
P500-060-030
19" x 2 HE
447 x 102 (88) x 541 (501) mm
10 kg
K347A SYSKON
P800-060-040
19" x 2 HE
447 x 102 (88) x 541 (501) mm
10 kg
K353A SYSKON
P1500-060-060
19" x 2 HE
447 x 102 (88) x 541 (501) mm
10 kg
K363A SYSKON
P3000-060-120
19" x 4 HE
447 x 191 (177) x 541 (501) mm
16 kg
K364A SYSKON
P4500-060-180
19" x 4 HE
447 x 191 (177) x 541 (501) mm
20 kg
K384A IEEE 488 interface