User Manual

GMC-I Messtechnik GmbH 35
8.7.1 Protective Conductor Current – IPE
I
PE
Single measurements, rotary switch level: green
Switch
Position
Measure-
ment Type,
With Mains
to Test
Socket
Measure-
ment Type,
Without
Mains to
Test Socket
Measuring Functions
I
PE
Direct
I
PE
Protective conductor current, RMS
I
PE~
AC component
I
PE=
DC component
U
LN
Test vol tage
Differential
I
PE
Protective conductor current, RMS
U
LN
Test vol tage
Alternative
I
PE
Protective conductor current, RMS
U
Test vol tage
AT3 adapter
1
I
PE
Protective conductor current, RMS
U
LN
Test vol tage
Clamp
2
I
PE
Protective conductor current, RMS
U
LN
Test vol tage
1
Adapter
AT3-I I I E,
AT3-I I S or AT3-I I S32:
Voltage measuring inputs for leakage
current measurement with differ-
ential method with SECUTEST PRO only (or instrument with feature I01)
2
Voltage measuring inputs for leakage current measurement with differ-
ential method using the WZ12C or SECUTEST CLIP current clamp sensor,
wi
th SECUTEST PRO only (or instrument with feature I01)
Applications
Protective conductor current must be measured for protection
class I devices.
Definition of Protective Conductor Current
(direct measurement)
Current which flows through the protective conductor in the case
of housings which are isolated from ground.
Definition of Differential Current
Sum of instantaneous current values which flow via the L and N
conductors at the device mains connection. Differential current is
practically identical to fault current in the event of an error. Fault
current: Current which is caused by an insulation defect, and
which flows via the defective point.
Definition of Alternative Measuring Method
(equivalent leakage current)
Equivalent leakage current is current which flows through the
active conductors of the device which are connected to each
other (L/N) to the protective conductor (SC1), or to the exposed,
conductive parts (SC2).
Differential Current Measuring Method
The device under test is operated with mains power. The sum of
the momentary values of all currents which flow through all active
conductors (L/N) at the mains side of the device connection is
measured. The measurements must be performed with mains
plug polarity in both directions.
Alternative Measuring Method (equivalent leakage current)
A high-impedance power supply is connected between the short-
circuited mains terminals and all exposed metal parts of the hous-
ing (which are connected to one another).
Current which flows over the insulation
at the device under test is
measured.
Protective Conductor Current Measuring Method
(direct measurement)
The device under test is operated with mains power. Current
which flows through the PE conductor to earth at the mains side
of the device connection is measured.
Note
Regardless of the currently selected connection type, all
help images and schematic diagrams can be queried for
the selected measuring function.
Direct Measuring Method
Dir
ect measurement type
– DUT mains plug to test socket
Schematic Diagram
The device under test is operated with mains power.
Protective conductor current is measured between the protective
conduct
or at the mains and the protective conductor terminal at
the DUT via the DUT’s mains cable.
Wiring Diagram
Differential current measurement
Differ
ential measurement type
– DUT mains plug to test socket
Schematic Diagram
The device under test is operated with mains power.
Differential current is measured between mains conductors L and
N (curr
ent clip concept).