User Manual

54 GMC-I Messtechnik GmbH
9 Special Functions – EXTRA
Depending on the device configuration, either the QR code for the
Internet link to the operating instructions or the measuring view for
the temperature measurement is displayed.
SECUTEST BASE(10)
EXTRA
Single measurements, rotary switch level: green
Switch
Position
Measuring Functions Measurement
Type,
EXTRA None None
QR code: Scanni
ng the QR code allows you to download and read
the current operating instructions from www.gossenme-
trawatt.com, for example at a tablet PC.
SECUTEST PRO (feature I01) and SECULIFE ST BASE(25)
EXTRA
Single measurements, rotary switch level: green
Switch
Position
Measuring Functions Measurement
Type
EXTRA Temperature V-COM
Clamp current V-COM
In this case, additional measuring fu
nctions are assigned to the
rotary switch’s EXTRA position.
S
elect the desired measuring function.
Measurement with Temperature Sensor
EXTRA
Temperature measurement functions with either a Pt100 or a
Pt1000 temperature sensor – the sensor type is automatically
detected internally.
Schematic Diagram
Wiring Diagram
Test Sequence with Temperature Sensor
Set the rotary switch to the EXTRA position.
Se
lect measuring function Temp.
Connec
t the temperature sensor’s plug to the V-COM sockets
at the test instrument.
Contac
t the device under test.
Start the test: press the START/STOP key.
The measured values are displayed. The measured
value recording symbol shown at the right appears.
Each time th
is key is pressed, the currently displayed
measured value is saved to the buffer.
S
top the test: press the START/STOP key.
The save symbol appears (floppy disk showing the
number of measured values
stored to buffer memory)
and prompts you to save the measured values to an ID
number.
P
ress the ESC key in order to discard the measured
values stored to buffer memory and acknowledge by
pressing the key shown at the right.