Meaning of Symbols in the User Interface – Test Sequence
Sym-
bol
Softkey Variants, Test Sequence
Test for Protection Class I Devices
Exposed, conductive parts are connected to the protec-
tive conductor so that they are
not charged with voltage if
the basic insulation should fail.
Test for Protection Class II Devices
These devices are equipped wit
h double insulation or
reinforced insulation.
Test for Protection Class III Devices
These devices are supplied with safety extra-low voltage
(SE
LV). Beyond this, no voltages are generated which
exceed SELV.
Type B applied parts (body)
Type BF applied parts (body float)
Type CF applied parts (cardiac float)
Configure sequence parameters (see page 61)
Set classification parameters
Assess visual inspection or function test with OK
or not OK (toggle key)
Enter a comment, e.g. for the visual inspection or function
test
Continue test, next test step in the test sequence
Stop continuous measurement, next test in test sequence
Accept changed parameter, return to memory view
Stop test sequence
– Repeat inspection (if it has been failed).
– Repeat test step
– Skip inspection test step
– Skip individual tests within
the test sequence
This option can be enabled for the
user in the setup under
„Auto Measurements“.
Start evaluation – record measured value. Each time this
softkey is pressed, an additional measured value is saved
and the number is increased by one.
Start evaluation sequence during a
continuous measurement
.
The number blinks.
Record measured value during the evaluation sequence of a
continuous measurement
.
Repeat measured value recording
Delete measured value
Display measured values
Display details from the results list
Hide details from the results list
The ID number to which the measurement(s) will be stored
can be entered here.
Valid measured values have been obtained for a test
sequence. This measurement can be saved.
Save measurement data as (with display of directory path
/ ID or new entry of an ID other than the preselected one)
Send measurement data to the PC, e. g. for storing in the
IZYTRON .IQ report generating softwar
e (push/print func-
tion), see IZYTRON .IQ online help for description.
10.2 Selecting and Configuring a Test Sequence
A1
Standard/
Connection
type
Protection class
Class parameter
Measurement type
Status line
Sequence parameters
Tree view*
Test sequence
Sample: Initial Page of a Test Sequence – Tree View
A1
Standard/
Connection Type Protection class
Class parameter
Measurement type
Status line
Sequence parameters
Applied parts
Detail view*
Test sequence
Sample: Initial Page of a Test Sequence – Detail View and
Applied Part
* SETUP switch position:
Setup menu 1/3 > Auto. Measurements > 2/2 > Initial
Window Style:
Tree or Detail View
Test Instruments with Feature E01 (Touch Screen)
You can switch between „Tree view“ and „Detail view“ (see above)
per „touch click“, that means by briefly tapping on the range of the
lower frame.
Classification Parameter – Automatic Detection
If the settings for certain classification parameters are automati-
cally detected by the test instrument, this is indicated respectively by
an or
ange frame (as of firmware V1.3.0; here: test socket connection
type and protection class I). Descriptions of these parameters are
listed in the following tables relative to the respective switch posi-
tions.
Automatic recognition for safety class active
When connecting or disconnecting a DUT, the
sa
fety class may be changed without request for
confirmation where necessary.
Automatic recognition for safety class active
The test instrument retains the se
lected safety class
setting when a DUT is connected or disconnected.