User Manual

GMC-I Messtechnik GmbH 59
Change Classification Parameters Conveniently
(Option Feature E01 (Touch Screen))
Per touch click (brief tapping) on the respective window of the
classification parameters the corresponing selection menu
opens.
Y
ou automatically go back to the start menu by selecting the
desired parameter.
Classification Parameter – VDE 0701-0702
Parameter Setting Options / Meaning
1/2
Standard/Test Sequence
VDE 0701-0702
VDE 0701-0702-IT, see following table
VDE 0701-0702-ExtC, see table below
VDE 0701-0702-PRCD, see table below
IEC 62353, see table below
IEC 60974-4, see table below
Protection class
1 2
Class I, Class II, Class I+II, Class I+III, Class II+III, Class I+II+III
Connection
1 2
test socket
Perman. connect.
Adapter: AT16/32-DI-Adap.
Adapter: VL2E-Adapter
Adapter: AT3-Adapter (feature I01)
Perman. Con.: P1+P2 (only with feature H01)
2/2
Meas. Mode
1)
passive
active
Auto detection Conn. & PC & MM
Conn. & PC
Conn. & MM
Conn. only
PC & MM
Prot. Class only
Meas. mode only
disabled:
all classification parameters such as connection and protection
class must be entered manually.
1
These parameters must be entered manually if they’re not automatically
detected, or if they’re detected incorrectly.
2
A limit value of the protective conductor resistance is determined on the
basis of length and cross-section. Data remain in memory until a new
entry is made.
Classification Parameter – VDE 0701-0702-IT
Parameter Setting Options / Meaning
1/2
Standard/Test Sequence VDE 0701-0702, see table above
VDE 0701-0702-IT
VDE 0701-0702-ExtC, see table below
VDE 0701-0702-PRCD, see table below
IEC 62353, see table below
IEC 60974-4, see table below
Protection class
1 2
Class I, Class II, Class I+II, Class I+III, Class II+III, Class I+II+III
Connection
1 2
test socket
Perman. connect.
Adapter: AT16/32-DI-Adap.
Adapter: AT3-Adapter
(feature I01)
2/2
Meas. Mode
1
active
Auto detection Conn. & PC & MM
Conn. & PC
Conn. & MM
Conn. only
PC & MM
Prot. Class only
Meas. mode only
disabled:
all classification parameters such as connection and protection
class must be entered manually.
1
These parameters must be entered manually if they’re not automatically
detected, or if they’re detected incorrectly.
2
A limit value of the protective conductor resistance is determined on the
basis of length and cross-section. Data remain in memory until a new
entry is made.
Classification Parameter – VDE 0701-0702-ExtC
Parameter Setting Options / Meaning
1/2
Standard/Test Sequence VDE 0701-0702, see table above
VDE 0701-0702-IT, see previous table
VDE 0701-0702-ExtC
VDE 0701-0702-PRCD, see following table
IEC 62353, see table below
IEC 60974-4, see table below
Protection class
1 2
Class I
Connection
1 2
test socket
Adapter: AT3-I I I E-Adapter
Adapter: EL1-Adapter
Adapter: VL2E-Adapter
2/2
Meas. Mode
1
ExtCord
2
Auto detection Conn. & PC & MM
Conn. & PC
Conn. & MM
Conn. only
PC & MM
Prot. Class only
Meas. mode only
disabled:
all classification parameters such as connection and protection
class must be entered manually.
1
These parameters must be entered manually if they’re not automatically
detected, or if they’re detected incorrectly.
2
A limit value of the protective conductor resistance is determined on the
basis of length and cross-section (with EL1 only length). Data remain in
memory until a new entry is made.