Momentary line voltage at the
SECUTEST test instrument is out-
side of the range permitted for a
10 A/25 A-R
PE
measurement (110
to 120 V or 220 to 240 V).
Th
e 10 A/25 A-R
PE
measurement
is only available when line voltage
is between 220 V and 240 V or
110 V and 120 V at 50 Hz or
60 Hz.
If you’r
e working with the SECUT-
EST in a system which does not lie
withi
n this voltage range, use one
of the 200 mA test currents in
order to determine protective con-
ductor resistance.
IT system option (see section 4.1.1
Measurements in IT Systems) is
activated.
An attempt has been made to
la
unch an active leakage current
measurement or a measurement
that refers to the PE at the mains
connection end (or a test sequence
which includes such measure-
ments).
Selec
t measurement type
"passive"
or
P
erform the requested tests in a
TT/TN system rather than in an IT
system and configure the SECUT-
EST for that purpose.
or
De
activate the leakage current
measurements in the sequence
parameters if possible.
Connection Error at the Test Socket
Test probe P1 is not connected.
Or
The test instrument’s 10 A/25 A
transformer is overheated.
Or
One of the fuses has blown (fuse
h
older in close proximity to the
mains input).
Repeat measurement with probe P1
connected.
Check the fuses and replace if nec-
essary.
Select a different test current (e.g.
200 mA) or wait until the trans-
former has cooled down and then
repeat the measurement.
Attention!
!
The 10 A/25 A measurement
is not suitable for continuous
operation!
A short-circuit has been detected
at the test socket between L and
N.
De
termine whether or not the de-
vice under test is defective.
In the case of DUTs which are intended
for operation at an outlet that’s pro-
tected with a 16 A fuse, a short-circuit
may be
detected under certain circum-
stances if, for example, they include a
PTC resistor (e.g. large floodlights).
Be sure to use a 3-phase test adapter
in order to test devices of this sort (e.g.
the AT3-I I I E).
You can skip over this short-circuit
message at your own risk and
place the device under test into
service. Any damage resulting from
skipping over this warning is ex-
cluded from the guarantee!
Error Messages Possible Causes Corrective Measures