User Manual

70 GMC-I Messtechnik GmbH
Database Processing Error
One of the fields was filled in with
invalid content while processing an
existing database object.
Please
be certain to complete all
mandatory fields (identified in red).
If necessary, check your entries to
the fields for invalid special charac-
ters.
The ID field was not filled in while
creating a new device.
Fill in the ID field.
There’s already an object with the
same ID under the “Customer”
database object.
An incorrect barcode has been
sel
ected.
Assign another ID.
Error while creating „.secu“ file on USB
flash drive
There is not or no longer enough free
spac
e on storage device left.
Particularly in the case of USB flash
dri
ves formatted for FAT16:
too many files on USB drive
Power consumption of USB flash
drive
in use exceeds 500 mA.
USB flash drive has been discon-
nected during the data import pro-
cess
USB flash drive is defective or
incompatibl
e with the SECUTEST.
Make sure that a minimum storage
capacity of 100 MB
is available on the
USB drive and/or delete any data files
no longer required.
If the problem persists, save the data
of the USB drive on another storage
device and format the USB drive (file
system FAT32).
Only use USB drives with a power
consumption of less than 500 mA
in combination with the SECUTEST
Mak
e sure that the USB drive is not
disconnected or moved until the
data export process is completed.
If none
of these measures leads to
an improvement, replace the USB
flash drive. A list of tested devices
is included in section 14.3.
Error Messages Possible Causes Corrective Measures