User Manual

GMC-I Messtechnik GmbH 71
Error while creating database backup file
on USB flash drive
There is not or no longer enough free
spac
e on storage device left.
Particularly in the case of USB flash
drives formatted for FAT16:
too many files on USB drive
Power consumption of USB flash
drive
in use exceeds 500 mA.
USB flash drive has been discon-
nected during the data import pro-
cess
USB flash drive is defective or
incompatibl
e with the SECUTEST.
Make sure that a minimum storage
capacity of 100 MB is available on the
USB drive and/or delete any data files
no longer required.
If the problem persists, save the data
of the USB drive on another storage
device and format the USB drive (file
system FAT32).
Only use USB drives with a power
consumption of less than 500 mA
in combination with the SECUTEST
Mak
e sure that the USB drive is not
disconnected or moved until the
data export process is completed.
If none
of these measures leads to
an improvement, replace the USB
flash drive. A list of tested devices
is included in section 14.3.
Moving of an object has failed
The moving of a test object would lead
to an ID conflict. The ID already exists
for this customer.
De
lete the object with double ID.
Select another customer as
removal target.
Errors during Operation with Barcode Scanner or RFID Scanner
The scanned barcode is too long.
While writing an RFID tag an
attempt was made to write an ID to
the tag with vowel mutations such
as ä, ü or ö, or with special charac-
ters.
Ch
ange vowel mutations such as ä
to ae.
A
void the use of special characters
in the ID.
Error Messages Possible Causes Corrective Measures