User Manual

GMC-I Messtechnik GmbH 77
1
Specified values are only valid for the display at the test instrument. Data
transmitted via the USB port may deviate from these values.
2
Known as equivalent leakage current or equivalent patient leakage current
from previous standards
3
Protective conductor current, touch current, device leakage current, pa-
tient leakage current
4
Protective conductor current, touch current, device leakage current
5
Only with
SECUTEST BASE10
(feature G01),
SECUTEST PRO
or
SECULIFE ST BASE
6
Only with SECUTEST PRO (feature I01) or SECULIFE ST BASE
7
Measurement types IPE_clamp and IG_clamp
8
Measurement type IPE_AT3 adapter and IG_AT3 adapter
9
The upper range limit depends on the selected test voltage.
10
Due to inrush current limiting components, the voltage at the test socket
may be lower than the measured line voltage.
11
SECULIFE ST BASE 25 only
Key: rdg. = reading (measured value), d = digit(s)
Test Times, Automated Sequence
Test times (“measurement duration” parameter) can be set separately
for each rotary switch position during configuration of the sequence
parameters. Test times are neither tested nor calibrated.
Emergency Shutdown During Leakage Current Measurement
As of 10 mA of differential current (can also be set to 30 mA),
automatic shutdown ensues within 100 ms. This shutdown does
not tak
e place during leakage current measurement with current
clamp sensor or adapter.
Influencing Quantities and Influence Error
Influencing Quantity /
Sphere of Influence
Designation
per
IEC 61557-16
Influence error
% r
dg.
Change of position E1
Change to test equipment supply
vol
tage
E2 2.5
Temperature fluctuation
E3
Specified influence error valid
start
ing with temperature changes
as of 10 K:
0 ... 40
C 2.5
Amount of current at DUT E4 2.5
Low frequency magnetic fields E5 2.5
DUT impedance E6 2.5
Capacitance during insulation
measurement
E7 2.5
Waveform of measured current
E8
49 51 Hz
2 with capacitive load (for equiva-
lent leakage current)
45 100 Hz 1 (for touch current)
2.5 for all other measuring ranges
I
Clamp
Current via
current clamp
sensor
1 mV : 1 mA]
(V–COM sockets
6 7
)
1 ... 99 mA
1 mA
(1 mV)
(2 % rdg.+2 d)
> 10 d
20 Hz ... 20 kH
z
without clamp
253 V Cont.
0.1 ... 0.99 A
0.01 A
(10 mV)
1.0 ... 9.9 A
0.1 A
(100 mV)
10 ... 300 A
1 A
(1 V)
Current via
current clamp
sensor
[10 mV : 1 mA]
(V–COM sockets
6 7
)
0.1 ... 9.9 mA
0.1 mA
(1 mV)
(2 % rdg.+2 d)
> 10 d
20 Hz ... 20 kH
z
without clamp
253 V Cont.
10 ... 99 mA
1 mA
(10 mV)
0.10 ... 0.99 A
0.01 A
(100 mV)
1.0 ... 30.0 A
0.1 A
(1 V)
Current via
current clamp
sensor
[100 mV : 1 mA]
(V–COM sockets
6 7
)
0.01 ... 0.99 mA
0.01 mA
(1 mV)
(2 % rdg.+2 d)
> 10 d
20 Hz ... 20 kH
z
without clamp
253 V Cont.
1.0 ... 9.9 mA
0.1 mA
(10 mV)
10 ... 99 mA
1 mA
(100 mV)
0.10 ... 3.00 A
0.01 A
(1 V)
Current via
current clamp
sensor
[1000 mV : 1 mA]
(V–COM sockets
6 7
)
1 ... 99 μA
1 μA
(1 mV)
(2 % rdg.+2 d)
> 10 d
20 Hz ... 20 kH
z
without clamp
253 V Cont.
0.10 ... 0.99 mA
0.01 mA
(10 mV)
1.0 ... 9.9 mA
0.1 mA
(100 mV)
10 ... 300 mA
1 mA
(1 V)
Func-
tion
Measured
Quantity
Display Range /
Nominal Range of
Use
Reso-
lution
Nominal
Voltage
U
N
Open-
Circuit
Voltage
U
0
Nom.
Current
I
N
Short-
Circuit
Current
I
K
Int.
Resis-
tance
R
I
Ref.
Resis-
tance
R
REF
Measuring
Uncertainty
1
Intrinsic Error
1
Overload
Capacity
Value Time