User Manual

80 GMC-I Messtechnik GmbH
This must be taken into consideration during technical safety
inspections or, instead of the insulation resistance measurement,
the protective conductor current measurement must result in a
value of less than 3.5 mA (or less than 7 mA if the equivalent leak-
age current method is used).
There are also 4 accessible conductive parts on the SECUT-
EST..., at which the touch current measur
ement must result in a
value of less than 0.5 mA:
Connector for service plug (jack socket)
•USB ports
Metallized start key
Protective conductor bar in the test socket
Note
In order to prevent damage to the SECUTEST... test
instrument, we recommend avoiding the performance of
measurements at the USB ports.
13.8 Returns and Environmentally Sound Disposal
The instrument is a category 9 product (monitoring and control
instrument) in accordance with ElektroG (German electrical and
electronic device law). This device is subject to the RoHS direc-
tive. Furthermore, we make reference to the fact that the current
sta
tus in this regard can be accessed on the Internet at
www.gossenmetrawatt.com by entering the search term WEEE.
We identify our electrical and electronic devices in
accordance with WEEE 2012/19/EU and Elektr
oG
using the symbol shown at the right per DIN EN 50419.
These devices may not be disposed of with the trash.
Please contact our service departmen
t regarding the return of old
devices (see address in Section 15).
14 Appendix
14.1 List of Suitable Printers with USB connection
The following devices have been tested for use with the test
instrument. We are unable to offer any guarantees regarding use
with other devices.
Z721S thermal printer
Z721D bar
code printer
(as of firmware V1.3.0, will be replaced by Z721E as from
2
018)
Setup options in the SETUP switch position
(Setup (2/3) > Printer > Z721
D > Printer settings):
Coding: Code 39, Code 128, EAN13, T
ext, QR Code, Micro
QR Code, DataMatrix
The respective paper siz
e will be adjusted automatically as
from FW 2.0 (6 mm, 9 mm, 12 mm, 18 mm, 24 mm, 36 mm)
Bar
code Printer Z721E (as from firmware V1.8.3)
Setting options in switch position SETUP
(Setup (2/3) > Printer > Z721E > Printer Settings)
Coding: Code39
, Code128, EAN13, Text, QR Code, Micro QR
Code, DataMatrix, Aztec
The respective paper size is adjusted automatically
(6 mm, 9 mm, 12 mm, 18 mm, 24 mm, 36 mm).
Note
Tapes
When using the label printer together with SECUTEST...,
only TZ(e)
tapes in widths of 6 mm, 9 mm, 12 mm,
18 mm, 24 mm and 36 mm are supported.
Note
2D Code Labels
For printing 2D code labels
(QR Code, MicroQR Code, DataMatrix, Aztec)
we recommend label cartridges with tape
widths of
12 mm or higher, however, at least 9 mm.
Note
Text Coding
In text print mode, print-out is limited to the CP1252
char
acter set – characters which are not representable,
are replaced by „_“.