Datasheet

GMC-I Messtechnik GmbH 3
SECUTEST BASE / PRO and SECULIFE ST BASE(25)
Test Instruments for
Measuring
Electrical Safety of Devices
Data Interfaces
Structures set up in, and measurement data saved to the test instru-
ment can be imported to
IZYTRONIQ
report generating software via the
USB slave port. Data can then be archived at the PC, comments can
be added with the software and reports can be generated.
The following input and output devices can be connected to the
two integrated USB master ports:
An external keyboard and a barcode or RFID reader,
USB stick for data backup
, import, export and reporting
•A printer
Software Update
The test instrument can always be kept current thanks to firm-
ware which can be updated via the USB slave port.
Report Generating Functions
All of the values required for approval reports or device logbooks for
electrical equipment (e.g. per ZVEH) can be measured with this
instrument. The measured data can be documented and archived
thanks to the measurement and test report that can be printed with a
thermal printer connected to the USB port, or stored to a PC.
Automatic Detection of Measuring Point Changes
During protective conductor measurement, the test instrument
recognizes whether or not the test probe is in contact with the
protective conductor, which is indicated by means of two different
acoustic signals. This function is very useful where several protec-
tive conductor connections need to be tested.
Mains Connection Analysis
Line voltage and frequency are measured and compared with the
data specified in the setup menu. Momentary voltage or nominal volt-
age in accordance with the standard is required, for instance in order
to extrapolate measured values for the leakage current measurement.
Automatic Detection of Mains Connection Errors
The device automatically recognizes mains connection errors if
the conditions in the following table have been fulfilled. The user is
informed of the type of error, and all measuring functions are dis-
abled in the event of danger.
1
10 A/25 A-R
PE
measurements are only possible with line voltages of 115/230 V
and line frequencies of 50/60 Hz.
2
if the test person is highly insulated, the following error message may appear: „In-
terference voltage at PE of mains connection
Analysis of DUT Connection and Condition
Depending on the measurement or how the DUT is connected, the fol-
lowing states are checked and displayed before measurement is begun.
*
applies to
M7050
with feature B00, B09 and B10
**
applies as from version 1.7.0; previous condition ≤ 1.5
Ω
or > 1.5
Ω
, respectively
Application
Regulations and standards in accordance with which the test instrument
is manufactured and tested:
Type of
Connection
Error
Message Condition Measurements
Voltage at protective
conductor PE to fin-
ger contact (START/
STOP key)
Display at the
instrument
Press START/STOP
button
U > 25 V
Button
PE:
< 1 MΩ
2
All measurements
disabled
Protective conductor
PE & phase conductor
L reversed and/or
neutral conductor N
interrupted
Voltage at PE
> 100 V
Impossible
(no supply power)
Line voltage
< 180 V / < 90 V
(depending on mains)
U
L-N
< 180 V
U
L-N
<90V
Possible under cer-
tain circumstances
1
Test on IT/TN system
Display at the
instrument
Connection
N PE > 50 kΩ
Possible under cer-
tain circumstances
Control Function Condition
Short-circuit test L–N Short-circuit / starting current R 2,5 Ω **
No short-circuit (AC test) R > 2,5 Ω **
Open-Circuit Voltage U
0
4.3 V, Short-Circuit Current I
K
< 250 mA
Short-circuit test N–PE Short-circuit R 2 kΩ
No short-circuit (AC test) R > 2 kΩ
Open-Circuit Voltage U
0
230 V, AC, Short-Circuit Current I
K
< 1.5 mA
On test On (passive DUT) R < 250 kΩ
Off (active DUT) R > 300 kΩ
Open-Circuit Voltage U
0
230 V AC, Short-Circuit Current I
K
< 1,5 mA
Special test No probe R > 2 MΩ
Probe detected R < 500 kΩ
Protection class detection
(only for country-specific (earth-contact) plug variant)*
Protective conductor exists: PC I R < 1 Ω
No protective conductor: PC II R > 10 Ω
Safety shutdown
Triggered at following residual current value (selectable)
> 10 mA / > 30 mA
Triggered at following residual current values (selectable)
During leakage current measurement > 10 mA
During protective conductor resistance meas. > 250 mA
Connection test
(only for country-specific
(earth-contact) plug variant)*
Checks whether the DUT is connected to the test socket.
Power line of DUT exists R < 1 Ω
No power line of DUT R > 10 Ω
Insulation test
DUT set up in a well-insulated fashion R 500 kΩ
DUT set up in a poorly insulated fashion R < 500 kΩ
PELine – PETestsocket:
Open-Circuit Voltage U
0
500 V DC, I
K
< 2 mA
Overcurrent protection (shutdown)
Shutdown in the event of a continuous flow of current via the test socket:
Our test instruments SECUTEST BASE(10), PRO and
SECULIFE ST
BASE(25)
allow for the active testing of devices with a nominal current (load
current) of up to 16 A. The test socket of the respective test instrument is
equipped with 16 A fuses and the switching capacity of the internal relays
also amounts to 16 A. Starting currents of up to 30 A are permissible. For
devices under test which are expected to feature a starting current of more
than 30 A, we strongly recommend the application of a test adapter for
higher starting currents: e. g. test adapter of the AT3 series
I > 16.5 A
DIN EN 61010-1:2011
VDE 0411-1:2011
Safety requirements for electrical equipment for measurement,
control and laboratory use – General requirements
DIN VDE 0404, part 1:
2002
Test and measuring equipment for testing the electrical safety of
electrical devices – General requirements
DIN VDE 0404, part 2:
2002
Equipment for testing after repairs and
modifications, or periodic testing
DIN VDE 0404, part 3:
2005
Equipment for periodic tests and tests prior to commission-
ing medical electrical devices or systems
DIN EN 60529/
VDE 0470, part 1
Test instruments and test procedures
Degrees of protection provided by enclosures (IP code)
DIN EN 61326-1
VDE 0843-20-1
Electrical equipment for measurement, control and laboratory
use – EMC requirements – Part 1: General requirements
IEC 61557-16 Electrical safety in distribution systems up to 1000 V a.c and
1500 V d.c –
Equipment for testing, measuring or monitoring of
protective measures - Part 16: Equipment for testing the safety
of electrical equipment and medical electrical equipment ac-
cording to IEC 62638 and IEC 62353 (IEC 85/437/CD:2012)