Calibration Data Report
Certificate number:
Q123987
Test Description
Supplied Value
Lower
limit
Measured
value
Upper
Limit
Status
TUR
<4
DC Volts
600mV Range
-
550 mV
-
550 mV
-
551.5
-
550.5
-
548.5
Pass
-
50 mV
-
50 mV
-
51.2
-
50.5
-
48.8
Pass
50 mV
50 mV
48.8
50.3
51.2
Pass
550 mV
550 mV
548.5
550.6
551.5
Pass
6V Range
-
5.5 V
-
5.5 V
-
5.
75
-
5.50
-
5.25
Pass
-
0.5V
-
0.5V
-
0.55
-
0.50
-
5.45
Pass
0.5V
0.5V
0.48
0.50
0.52
Pass
3V
3V
2.98
3.01
3.02
Pass
5.5V
5.5V
5.48
5.50
5.52
Pass
AC Volts
600mV Range
50 mV
@53Hz
50 mV
@ 53Hz
48.5
49.8
51.5
Pass
550mV
@ 53Hz
550mV
@ 53Hz
548.5
550.2
551.5
Pass
6V Range
1 V@ 10Hz
1 V@ 10Hz
0.98
1.00
1.02
Pass
1V @ 53Hz
1V @ 53Hz
0.99
1.00
1.01
Pass
1V @ 1KHz
1V @ 1KHz
0.98
0.99
1.02
Pass
1V @ 5KHz
1V @ 5KHz
0.95
0.98
1.05
Pass
Resistance
600
Ω Range
60
60
58.5
59.0
61.5
Pass
550
550
548.5
549.0
551.5
Pass
6K
Ω Range
1K
1K
0.975
1.00
1.025
Pass
5.5K
5.5K
5.400
5.50
5.600
Pass