6.1 Temperature Simulation for Resistance Temperature Detectors –
RTD setting for temperature
Resistance temperature detectors (types Pt100, Pt1000, Ni100
and Ni1000) are simulated by means of resistance values.
Abbreviated Instructions
Select the calibration function.
Select a sensor value and acknowledge for the fixed value function.
select sensor Pt100 ...ni1000
Set the temperature simulator value.
120.0°C  
Response time of the calibrator output to the specified resistance
value after measuring current is applied is max. 30 ms.
Devices under test with non-continuous measuring current (e.g.
scanned measuring inputs) result in erroneous measured values if
measurement is started before response time has elapsed. The
calibrator cannot be used for devices of this type.
6.2 Temperature Simulation for Thermocouples – TC Temperature Setting
Thermocouples (types B, E, J, K, L, N, R, S, T and U) are
simulated with voltage. Internal or external temperature
compensation is possible.
Abbreviated Instructions
Select the calibration function.
Select a sensor type and acknowledge for the fixed value function.
select sensor b ...u
Set the temperature simulator value.
120.0 °C  
Select internal or external reference temperature, set external reference
temperature (see page 11).
RTD
ON/OFF
LIGHT
SELECT
ESC
OUT
ENTER
TC
ON/OFF
LIGHT
SELECT
ESC
OUT