Setting the Interval Parameters
seleCt range 300 mV 15 V 1nt
Initial value:
1nt start 02.000 V  
Final value: 1nt end 10.000 V  
Steps:
1nt steps 03.0  
Dwell time:
1nt t1 00.05 min.s  
Repetition:
1nt mode auto manual
(Auto = automatic sequence, MAnuAL = manual sequence)
Manually Controlled Interval Sequence
After entering all parameters for the “manual interval sequence”
output function (
1nt mode = manual ) and starting the function with
the key, the individual steps are triggered with the and
keys. The relationship between the output signal and each of
the key operations is depicted with the help of the following exam-
ple.
Example of a Manually Controlled Interval Sequence
Key
1 When 1nt
ready is displayed:
Start the sequence by pressing .
2 The sequence is continued in the corresponding direction by
pressing the or the key.
3 Stop the interval sequence by pressing .