User Manual

18 GMC-I Messtechnik GmbH
Automatic Interval Sequence
Automatic execution of a programmed sequence range is above
all advisable if feeding to a signal circuit and scanning of the
peripheral device under test are physically separated.
After entering all parameters for the “automatic interval sequence”
output function (see above) (I
nt, mode = auto), the sequence can be
started and stopped whenever desired, as well as resumed.
Example of an Automatic Interval Sequence
Interval parameters: output quantity: U (range of 0 15 V),
start =2V, end = 10 V, number of interval steps =3, t1 =5s,
mode = auto
Key
1 When 1nt
ready is displayed:
Start the sequence by pressing .
2 The sequence is stopped by pressing the or the key.
Interval time elapsed thus far is saved as t
x
.
3 The sequence is resumed by pressing the key; remaining
sequence duration t
y
=t1–t
x
.
4 Stop the interval sequence by pressing .
2
4.6
7.3
10
t[s]
U
[V]
run up stop
➀➃
5101520250
run dn hold run up run dn
HOLD CONT
LCD:
ESC/
OUT
ENTER
SELECT
ESC