Setting Ramp Parameters
seleCt range 300 mV 15 V ramp
Initial value:
ramp start 02.000 V  
Final value: ramp end 10.000 V  
Rise time:
ramp t1 00.05 min.s  
Dwell time:
ramp t2 00.08 min.s  
Decay time:
ramp t3 00.05 min.s  
Repetition: ramp mode repeat onCe
(rEPEAT = endless loop, onCE = one-time only)
Example of a Periodic Ramp Sequence
Manually Controlled Ramp Sequence
After entering all parameters, start by pressing .
Rising or decaying ramps can be triggered with the or key.
The relationship between the output signal and each of the key
operations is depicted with the help of the following example.
Example of a Periodic Ramp Sequence Controlled by Manual Intervention
Ramp parameters: output quantity: U (range of 0 15 V),
start =2V, end =10V, t1 =5s, t2 =8s, t3 =5s, repeat for peri-
odic ramps
Key
1 When ramp ready is displayed:
Start the sequence by pressing .
2 Stop the decaying ramp within decay time t3 with the or
key.
3 Start a rising ramp within remaining decay time t3 with the
key.
4 Stop the ramp sequence by pressing the or key.
5 Start a decaying ramp with the key, remaining dwell time t2
is cleared.
6 Stop the ramp sequence by pressing .
SELECT
ESC
CAL
SETUP
OUT
ENTER
OUT
ENTER
OUT
ENTER
OUT
ENTER
OUT
ENTER
Start
End
t1
t2
t3
t2
Output
t
2 V
10 V
t1 t2 t3
LCD:
t3 t2
t1
➀➁
t1 up t2 run t3 dn t3 hld t3 up t3 run t3 hld t1 dn