User Manual

8 GMC-I Messtechnik GmbH
3 Voltage Source [V]
Voltages can be simulated within the following ranges:
0 60 mV, 0 300 mV, 0 3 V, 0 10 V and 0 15 V.
Resistance of the connected circuit should not be any less
than 1 k.
Connect the device to be calibrated with the measurement
cable as shown.
Select the V calibration function with the rotary switch.
Switch the calibrator on by pressing the ON / OFF | LIGHT
key.
The last selected voltage range is displayed.
Set the voltage value:
ON indicates:
Voltage is applied directly to the output!
Select the digit to be changed with the keys, and change
the respectively selected digit with the keys.
The output can be deactivated with the OUT | ENTER key
[
out.off], or activated once again.
Selecting a Voltage Range for the Fixed Value Function
Press the SELECT | ESC key in order to switch to the
[select range] menu.
Select the desired voltage range with the  keys.
Acknowledge your selection with OUT | ENTER.
The display is switched to the voltage value entry window, but
the selected voltage range still appears in the auxiliary display.
Selecting the Voltage Range for the Interval or Ramp Function
Press the SELECT | ESC key in order to switch to the
[
select range] menu. Select the desired voltage range with the
 keys.
Switch to the interval or ramp function menu with the 
keys (see section 8). Start the respective function by pressing
OUT | ENTER.
Abbreviated Instructions
Select calibration function.
Select voltage range and acknowledge for fixed value function.
select range: 15 V 60 mV
Change the fixed value.
000.00 V  
(Negative values within a range of 60 mV or 300 mV can be selected by
scrolling with the
key to below the zero point.)
Output
CALIBRATOR
+
+
Input
Device to be calibrated, e.g.
measuring transducer
1k
V
ON/OFF
LIGHT
SELECT
ESC