User Manual

GOSSEN METRAWATT GMBH 63
1)
nur für Netze mit Überspannungskategorie II, Verschmutzungsgrad 2, max. 5 min
2)
U > 253 V nur mit 2-Pol-Adapter
3)
100 U
N
x1/
4)
mit externem Adapter Z541A als Zubehör
5)
I
N
= 500 mA, max. U
N
= 330 V
6)
L-PE: 300 V, L-L: 500 V
Z
Schl
Z
I
Z
Schl
(Vollwellen)
Z
I
0,01 ... 9,99 10 m 0,83 ... 4,0 A
0,15 ... 0,49
0,50 ... 0,99
1,00 ... 9,99
U
N
= 120/230 V
U
N
2)
= 400 V/
500 V bei Z
Schl
f
N
= 50/60 Hz
±(10% v.M.+2D)
±(10% v.M.+3D)
±(5% v.M.+3D)
±5 D
±(4% v.M.+3D)
±(3% v.M.+3D)
Z
Schl
Z
Schl
(+/– Halbwellen)
0,25 ... 0,99
1,00 ... 9,99
±(20% v.M.+5D)
±(10% v.M.+3D)
±(6% v.M.+5D)
±(4% v.M.+3D)
I
K
0 A ... 999 A
1,00 kA ... 9,99 kA
10,0 kA ... 50,0 kA
3)
1 A
10 A
100 A
120 (108 ... 132)
V
230 (196 ... 253)
V
400 (340 ... 440)
V
——
R
E
R
E
(R
ESchl
ohne
Sonde)
0 ... 10
0 ... 10
0 ... 10
0 ... 100
0 ... 1 k
1 k ... 10 k
10 m
10 m
10 m
10 m
1
1
0,83 ... 3,4 A
0,83 ... 3,4 A
0,83 ... 3,4 A
400 mA
40 mA
4 mA
0,15 ... 0,49
0,50 ... 0,99
1,0 ...9,99
10 ...99,9
100 ...999
1 k ...9,99 k
U
N
= 120/230 V
U
N
= 400 V
2)
f
N
= 50/60 Hz
±(10% v.M.+2D)
±(10% v.M.+3D)
±(5% v.M.+3D)
±(10% v.M.+3D)
±(10% v.M.+3D)
±(10% v.M.+3D)
±5 D
±(4% v.M.+3D)
±(3% v.M.+3D)
±(3% v.M.+3D)
±(3% v.M.+3D)
±(3% v.M.+3D)
●●
U
E
0 ... 253 V 1 V Rechenwert
Z
ST
0 ... 1 M 1 k
2,3 mA
bei 230 V
10 k ... 199 k
200 k
... 999 k
U
0
= U
L-N
±(10% v.M.+2D)
±(20% v.M.+2D)
±(5% v.M.+3D)
±(10%
v.M.+3D)
R
ST
10 k ... 199 kΩ±(20% v.M.+2D)
±(20%
v.M.+3D)
R
ISO
R
ISO
, R
E(ISO)
0,01 ... 9,99 M
10,0 ... 99,9 M
10 k
100 k
I
K
= 1,5 mA 50 k... 100 M
U
N
= 100 V
I
N
= 1 mA
±(5% v.M.+1D) ±(3% v.M.+1D)
●●
0,01 ... 9,99 M
10,0 ... 99,9 M
100 ... 200 M
10 k
100 k
1 M
U
N
= 250 V
I
N
= 1 mA
0,01 ... 9,99 M
10,0 ... 99,9 M
100 ... 300 M
10 k
100 k
1 M
U
N
= 500 V
I
N
= 1 mA
U 25 ... 600 V– 1 V 500 k 25 ... 600 V ±(3% v.M.+1D)
±(1,5%
v.M.+1D)
R
LO
R
LO
0,01 ... 9,99
10,0 ... 99,9
10 m
100 m
I
m
200 mA 0,1 ... 6 U
0
= 4,5 V ±(4% v.M.+2D) ±(2% v.M.+2D)
Funk-
tion
Messgröße Messbereich
(Anzeigebereich I
K
)
Auf-
lösung
Eingangs-
impedanz/
Prüfstrom
Nenngebrauchs-
bereich
Nennwerte Betriebsmess-
abweichung
Eigen-
abweichung
Anschlüsse
Stecker-
einsatz
2)
2-Pol-
Adapter
3-Pol-
Adapter
Sonde Zange
Z541
A