User manual

GMC-I Messtechnik GmbH 9
Electrical Safety
Safety class I
Overvoltage
category II for mains input
I for output and interfaces
Fouling factor 2
Earth leakage
current 2.5 mA, typ.
IEC 61010-1:1990 + A1:1992 / DIN EN 61010-1: 1993 /
VDE 0411-1:1994
DIN VDE 0160:1988 + A1:1989 class W1
EN 60950:1992 / VDE 0805:1990
Protection IP 20 for housing per
IEC 529:1989,
EN 60529:1991,
VDE 0470-1:1992
Electrical isolation Test voltage
Mains/output–PE 1.35 kV~
Mains–output 2.7 kV~ (type test: 3.7 kV ~)
Electromagnetic Compatibility (EMC)
Product standard EN 61326-1:1997 + A1: 1998
Interference
emission EN 55022:1998 class A
Interference
immunity EN 61000-4-2:1995, feature A
EN 61000-4-3:1996 + A1:1998,
feature B
EN 61000-4-4:1995, feature B
EN 61000-4-5:1995, feature B
EN 61000-4-6:1996, feature B
EN 61000-4-11:1994, feature B
Ambient Conditions
IEC 68-2-6 (’90) Vibration resistance
10 ... 55 Hz, 0.3 mm, 1 octave / min.,
3 x 30 min.
IEC 68-2-27 (’89) Impact resistance
(15 g, 11 ms, semi-sinusoidal,
3 x 6 shocks)
Temperature Operation: 0 to 50° C,
range current derating at > 40° C
(see also chapter 1.5.3)
Storage: -25 to +75° C
Relative humidity Operation:
75%, no condensation allowed
Cooling With integrated fan
Inlet vent: side panel
Outlet vent: rear panel
1.5.2 Mechanical Data
Type Benchtop device, suitable for rack
mounting
Dimensions See also dimensional drawing in
Figure 1.5
(W x H x D) Benchtop device:
221.5 x 102 x 397.5 mm
For 19" rack: ½19" x 2 standard height
units x 400 mm
Weight Approx. 2.8 kg
IEEE 488 interface (optional):
approx. 0.1 kg