750000 ALTERNATING PRESSURE / LOW AIR LOSS MATTRESS SYSTEM USER MANUAL Important: Do not operate the Mattress System without first reading and understanding this manual! Save this manual for future use. Info: The most current version of this manual can be found online at www.grahamfield.com.
CONTENTS INTRODUCTION ..................................................................................... 3 INTENDED USE OF THIS DEVICE ........................................................ 3 CONTRAINDICATION...................................................................... 3 IMPORTANT SAFETY PRECAUTIONS .................................................. 4 DANGER ................................................................................................. 4 WARNING ..........................
INTRODUCTION Use this manual for set-up and operation of the Lumex AltaDyne 750000 Alternating Pressure / Low Air Loss Mattress System. Read all instructions before using the mattress system. Save this manual for future reference. INTENDED USE OF THIS DEVICE The intended use of the Lumex AltaDyne 750000 Alternating Pressure / Low Air Loss Mattress System is: • The treatment of pressure ulcers while optimizing user comfort. • Pain management as prescribed by a physician.
IMPORTANT SAFETY PRECAUTIONS The safety statements presented in this chapter refer to the basic safety information that should be observed by those using the Alternating Pressure / Low Air Loss Mattress System. There are additional safety statements in other chapters or sections, which may be the same as or similar to the following, or specific to the operations. DANGER: Indicates an imminent hazard situation that, if not avoided, will result in death or serious injury.
WARNING To reduce the risk of burns, electrocution, fire, or personal injury: WARNING: Do not leave this product unattended when plugged in. WARNING: Always use close supervision when this product is used by, on, or near children or those who require close supervision. WARNING: Use this product only as intended and described in this manual. Do not use attachments or accessories not recommended by Graham-Field. WARNING: Never operate this product if: a) It has a damaged cord or plug.
WARNING: Patient entrapment with bed side rails may cause injury or death. The bed frame and its components, including the mattress, bed side rails, head and foot board, bedding, and any accessories added to the bed, can all affect the risk of entrapment. Thorough patient assessment and monitoring are necessary to reduce the risk of entrapment, including establishing whether the use of a bed rail is in the best interest of the patient. Read and understand the User Manual before using this equipment.
EQUIPMENT SYMBOLS Attention! Consult accompanying documents (this manual) “BF”symbol: Indicates that this product is in accordance with the degree of protection against electric shock for type BF equipment Functional earth (for UL only) Class II (for UL only) Disposal of Electrical & Electronic Equipment (WEEE): Do not treat this product as household waste.
PRODUCT DESCRIPTION PUMP AND MATTRESS SYSTEM foot end symbol CPR valves mattress pump mattress-pump connectors pump and mattress system Pump: front quick connector 8 750000-INS-LAB-RevC11
Pump: rear Pump: top panel static function switch main power switch low pressure indicator normal pressure indicator pressure adjustment knob Pressure adjust knob: The pressure adjust knob controls the air pressure output. Turn it clockwise to increase output pressure (firmer); turn it counter-clockwise to decrease air output pressure (softer). Higher pressure output will better support the heavier-weight user. The setting is measured in kilograms (kg).
Static function switch: Press the static function switch to suspend alternating function, if needed; the pressure inside the air tubes will then be adjusted to identical pressure. Press the static function switch again to change back to alternating mode. The static function switch toggles between static and alternating modes. Main power switch: The main power switch toggles between ON and OFF. When power is ON, either the normal or low pressure indicator illuminates.
INSTALLATION UNPACKING 1. Before unpacking, check for obvious damage to the carton or its contents. If damage is evident, please notify the carrier and your Graham-Field equipment provider. 2. Remove all loose packing from the carton. 3. Carefully remove all components from the carton. 4. Inspect all components. If damage is evident, please notify the carrier and your Graham-Field equipment provider.
PUMP AND MATTRESS INSTALLATION 1. Place the mattress on top of the bed frame; ensure that the foot end symbol faces up at foot end as shown at right. 2. Fold open the wire hangers on the back of the pump as shown at right. 3. Either hang the pump on the bed's foot end as shown at right, or place the pump on a flat surface easily accessible to the user, caregiver, and/or doctor. s NOTICE: Place the device in a position where the user, caregiver, and/or doctor can access it easily. 4.
WARNING: Locate all cords so that they will not be stepped on, tripped over, or otherwise subjected to damage or stress. WARNING: Do not use a household extension cord if the electrical cord does not reach the power outlet. Use of an improper extension cord could result in fire and electric shock. If an extension cord must be used, use a three conductor cord with ground, properly wired, in good electrical condition, and keep it as short as possible.
OPERATION 1. Turn the power switch to the ON position. 2. Adjust the pressure adjust knob to the maximum position. Info: Set the pressure adjust knob to the maximum position for the quickest inflation at set-up; the mattress can then be adjusted to the desired pressure. 3. The pump will begin delivering air into the mattress. While the mattress is filling, the low pressure indicator (yellow LED) will illuminate. 4.
under the patient's buttock area; your fingers should just fit. If the fit is too loose, decrease the setting. If your fingers do not fit well, increase the setting two notches firmer. Wait ten minutes and then check again. You should only have to adjust this setting one time per patient. Please consult your physician for an appropriate setting. QUICK DISCONNECT Pull the power plug from the wall connector to disconnect the device quickly.
MAINTENANCE CLEANING Perform the following cleaning procedures before use. Pump: Wipe the pump with a clean, damp cloth and mild detergent, and keep it away from dust. Air dry. s NOTICE: Do not use phenolic products or corrosive or powdered cleansers to clean the pump. s NOTICE: Do not immerse or soak the pump. Mattress: Wipe the mattress unit with a clean, damp cloth and mild detergent. The mattress may also be cleaned using a 10% solution of sodium hypochlorite (bleach) diluted in water.
Fuse replacement tool needed: small screwdriver 1. If you suspect a blown fuse, disconnect the plug from the wall outlet immediately. 2. Use a small screwdriver to remove the cover of the fuse holder (fuse locations are shown in picture at right). 3. Insert a new fuse of the correct rating (T1A/250V, VDE approved). 4. Replace the fuse holder cover. Air filter replacement 1. Remove the air filter plate located at rear of pump (filter location shown in picture at right). 2. Remove the filter.
SPECIFICATIONS Pump Specification Power Supply 120V System: AC 120V 60 Hz Info: see rating label on product Fuse Rating T1A, 250V Cycle time 8 minutes Air input >4 LPM Dimensions (L x W x H) 9.8" x 5.3" x 3.7" (25 x 13.5 x 9.5 cm) Weight 4 lb (1.
WARRANTY GF Health Products, Inc. warrants the Lumex AltaDyne 750000 Alternating Pressure / Low Air Loss Mattress System for a period of twelve months for defects in workmanship and materials. If a product is deemed to be under warranty, GF Health Products, Inc. shall provide, at its option, (1) replacement of any defective part or product or (2) a credit of the original selling price made to GF Health Products, Inc.’s initial customer.