Manual
Table Of Contents
- INTERFACE OVERVIEW
- COMMAND OVERVIEW
- COMMAND DETAILS
- Common Commands
- Acquisition Commands
- Autoscale Commands
- Vertical Commands
- Math Commands
- Cursor Commands
- Display Commands
- Hardcopy Commands
- Measure Commands
- Measurement Commands
- Reference Commands
- Run Command
- Stop Command
- Single Command
- Force Command
- Timebase Commands
- Trigger Commands
- System Commands
- Save/Recall Commands
- Ethernet Commands
- Time Commands
- Bus Decode Commands
- Mark Commands
- Search Commands
- Digital Commands
- Label Commands
- Utility Commands
- APPENDX
- INDEX
COMMAND DETAILS
105
Syntax
:MEASUrement:STATIstics:MODe {OFF | ON | ?}
Related
commands
:MEASUrement:STATIstics
Parameter/
Return parameter
ON
Display the statistics on the screen.
OFF
Remove the statistics from the screen
Example
:MEASUrement:STATIstics:MODe ON
Displays statistics on the screen.
:MEASUrement:STATIstics:WEIghting
Set
Query
Description
Sets and queries the number of samples used for
the statistics calculations.
Syntax
:MEASUrement:STATIstics:WEIghting { <NR1> | ? }
Parameter/
Return parameter
<NR1>
Number of samples (2~1000)
Example
:MEASUrement:STATIstics:WEIghting 5
Sets the number of samples to 5.
:MEASUrement:STATIstics
Set
Description
Resets the statics calculations. This command will
clear all the currently accumulated measurements.
Syntax
:MEASUrement:STATIstics