Manual
Table Of Contents
- INTERFACE OVERVIEW
- COMMAND OVERVIEW
- COMMAND DETAILS
- Common Commands
- Acquisition Commands
- Autoscale Commands
- Vertical Commands
- Math Commands
- Cursor Commands
- Display Commands
- Hardcopy Commands
- Measure Commands
- Measurement Commands
- Reference Commands
- Run Command
- Stop Command
- Single Command
- Force Command
- Timebase Commands
- Trigger Commands
- System Commands
- Save/Recall Commands
- Ethernet Commands
- Time Commands
- Bus Decode Commands
- Mark Commands
- Search Commands
- Digital Commands
- Label Commands
- Utility Commands
- APPENDX
- INDEX
GDS-2000A Programming Manual
122
Related
commands
:TRIGger:DELay:TYPe
Parameter
<NRf>
Delay time (1.00E-8~1.00E+1)
Return parameter
<NR3>
Returns the delay time.
Example
:TRIGger:DELay:TIMe 1.00E-6
Sets the delay time to 1us.
:TRIGger:DELay:EVENt
Set
Query
Description
Sets or queries the number of events for the event
delay trigger.
Syntax
:TRIGger:DELay:EVENt {<NR1> | ?}
Related
commands
:TRIGger:DELay:TYPe
Parameter
<NR1>
1~65535 events
Return parameter
<NR1>
Returns the number of events.
Example
:TRIGger:DELay:EVENt 2
Sets the number of events to 2.
:TRIGger:DELay:LEVel
Set
Query
Description
Sets or queries the trigger delay level.
Syntax
:TRIGger:DELay:LEVel {<NRf> | ?}
Parameter
<NRf>
Delay trigger level
Return parameter
<NR3>
Returns the delay trigger.
Example
:TRIGger:DELay:LEVel 5.00E-3
Sets the delay trigger to 5mV/mA.
:TRIGger:PULSEWidth:POLarity
Set
Query
Description
Sets or queries the pulse width trigger polarity.