Manual
Table Of Contents
- INTERFACE OVERVIEW
- COMMAND OVERVIEW
- COMMAND DETAILS
- Common Commands
- Acquisition Commands
- Autoscale Commands
- Vertical Commands
- Math Commands
- Cursor Commands
- Display Commands
- Hardcopy Commands
- Measure Commands
- Measurement Commands
- Reference Commands
- Run Command
- Stop Command
- Single Command
- Force Command
- Timebase Commands
- Trigger Commands
- System Commands
- Save/Recall Commands
- Ethernet Commands
- Time Commands
- Bus Decode Commands
- Mark Commands
- Search Commands
- Digital Commands
- Label Commands
- Utility Commands
- APPENDX
- INDEX
COMMAND DETAILS
123
Syntax
:TRIGger:PULSEWidth:POLarity
{POSitive | NEGative | ?}
Related
commands
:TRIGger:TYPe
Parameter
POSitive
Positive polarity
NEGative
Negative polarity
Return parameter
Returns the pulse width polarity.
Example
:TRIGger:PULSEWidth:POLarity POSitive
Sets the pulse width polarity to positive.
:TRIGger:RUNT:POLarity
Set
Query
Description
Sets or queries the Pulse Runt trigger polarity.
Syntax
:TRIGger:RUNT:POLarity { POSitive | NEGative |
EITher | ? }
Related
commands
:TRIGger:TYPe
Parameter
POSitive
Positive polarity
NEGative
Negative polarity
EITher
Positive or negative polarity
Return parameter
Returns the pulse runt trigger polarity.
Example
:TRIGger:RUNT:POLarity POSitive
Sets the Pulse Runt trigger polarity to positive.
:TRIGger:RUNT:WHEn
Set
Query
Description
Sets or queries the Pulse Runt trigger conditions.
Syntax
:TRIGger:RUNT:WHEn {THAN | LESSthan | EQual |
UNEQual | ? }
Related
commands
:TRIGger:TYPe
:TRIGger:RUNT:TIMe