Manual

COMMAND DETAILS
143
Parameter
<X>
X is the digital channel number
D(0~15)
HIGH
Set to a high logic state
LOW
Set to a low logic state
X
Set to a "don’t care" state.
Return parameter
Returns the logic state of the selected channel (HIGH,
LOW, X).
Example1
:TRIGger:LOGic:PATtern:INPut:D0?
X
:TRIGger:LOGic:PATtern:DELTatime
Set
Query
Description
Sets or returns the pattern trigger delta time value.
Syntax
:TRIGger:LOGic:PATtern:DELTatime {<NR3> | ? }
Related
commands
:TRIGger:LOGic:PATtern:WHEn
Parameter/Return
parameter
<NR3>
Pattern trigger delta time in seconds.
(10nS to 10S)
Example
:TRIGger:LOGic:PATtern:DELTatime?
8.960e-05
:TRIGger:LOGic:PATtern:WHEn
Set
Query
Description
Sets or returns the pattern logic condition on
which to trigger the oscilloscope.
Syntax
:TRIGger:LOGic:PATtern:WHEn { TRUE | FALSE |
LESSthan | THAN | Equal | UNEQual | ? }
Related
commands
:TRIGger:LOGic:PATtern:DELTatime
Parameter/
Return parameter
TRUE
Set true mode.
FALSE
Set false mode.