Manual

COMMAND DETAILS
183
:SEARCH:TRIGger:BUS:B1:I2C:ADDRess:V
ALue
Set
Query
Description
Sets or queries the I
2
C bus address value when the
I
2
C search trigger is set to trigger on Address or
Address/Data.
Syntax
:SEARCH:TRIGger:BUS:B1:I2C:ADDRess:VALue
{string | ? }
Related
commands
:SEARCH:TRIGger:BUS:B1:I2C:ADDRess:MODe
Parameter
<sting>
7/10 characters, must be enclosed in
double quotes "string".
x = don’t care
1 = binary 1
0 = binary 0
Return Parameter
Returns the address value in binary.
Example1
:SEARCH:TRIGger:BUS:B1:I2C:ADDRess:MODe
ADDR7
:SEARCH:TRIGger:BUS:B1:I2C:ADDRess:VALue
"xxx0101"
Sets the address to XXX0101
Example 2
:SEARCH:TRIGger:BUS:B1:I2C:ADDRess:VALue?
XXX0101
:SEARCH:TRIGger:BUS:B1:I2C:ADDRess:DI
Rection
Set
Query
Description
Sets or queries the address bit as read write or
don’t care for the search function.
Note
This setting only applies when the I
2
C search
trigger is set to trigger on Address or
Address/Data