GDS-2000A Programming Manual
78
:MEASure:METHod
Set
Query
Description
Sets or queries the method used to determine the
High-Low measurement values.
Syntax
:MEASure:METHod { AUTo | HIStogram | MINMax
| ? }
Parameter
AUTo
Set to auto.
HIStogram
Set to the Histogram method.
MINMax
Set to the Min-Max method.
Return parameter
Returns the measurement method (AUTO,
HISTOGRAM, MINMAX)
Example
:MEASure:METHod: AUTo
Set the measurement method to auto.
:MEASure:FALL
Query
Description
Returns the fall time measurement result.
Syntax
:MEASure:FALL{?}
Related
Commands
:MEASure:SOURce<X>
Return parameter
<NR3>
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:FALL?
Selects Channel 1 as the source, and then measures
the fall time.