Manual

GDS-2000A Programming Manual
92
Related
Commands
:MEASure:SOURce<X>
Return parameter
<NR3>
Returns the area value.
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:AREa?
1.958E-03
Selects Channel 1, and then measures the area.
:MEASure:CARea
Query
Description
Returns the voltage/current area over one full
period.
Syntax
:MEASure:CARea{?}
Related
Commands
:MEASure:SOURce<X>
Return parameter
<NR3>
Returns the area value.
Chan Off
Indicates the source channel is not
activated.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:CARea?
1.958E-03
Selects Channel 1, and then measures the area.