Datasheet

GW Instek GPM-8310 is a digital power meter for single-phase (1P/2W) AC power measurement. Features include DC, 0.1Hz~100kHz test
bandwidth, 16bits A/D, and 300 kHz sampling rate. It adopts 5” TFT LCD screen with a five-digit measurement display and provides 25
power measurement related parameters, and has a high-precision measurement capability. It also features the ability to display waveform
(voltage/current/power), the integration measurement function, harmonic measurement and analysis of each order (meeting the IEC
61000-4-7 harmonics measurement requirements at 50/60Hz), external sensor input terminals, and various communication interfaces,
etc., to help users achieve clear, convenient and accurate power measurements. This power meter is a most cost-effective power meter
with most complete functionalities among the products of the same category.
The rated direct input voltage of GPM-8310 is 600V and the input current is 20A. The minimum current level is 5mA (resolution up to
0.1uA) and the power measurement resolution is 0.1uW. The crest factor can reach 3 (half measurement range can reach 6 or 6A), and the
voltage/current/power measurement capability can reach (±0.05% reading ±0.1% level). Different measurement modes can be selected
according to ( AC+DC/ AC/ DC/ V-MEAN), providing up to 25 relevant parameters for power measurement, including voltage (Vrms/ Vac/
Vdc/ Vmn/ V+pk / V-pk), current (Irms/ Iac/ Idc/ I+pk/ I-pk), frequency (VHz/ IHz), power (P/ P+pk/ P-pk), crest factor (CFV/ CFI),
apparent power (VA), reactive power (VAR), power factor (PF), phase angle (DEG), total harmonic distortion rate (THDV/THDI),
maximum current ratio (MCR), and the MATH calculation function. Hence, for the measurement of low current/low power such as
standby power consumption, or the measurement of power consumption of general products, this power meter provides the best range
and accuracy support.
GPM-8310 also makes good use of the advantages of the TFT LCD to display the results of parameter measurement by using numerical
and graphical methods. In terms of numerical values, the general mode and the simple mode are provided. The general mode can display
10 measurement parameters (2 main measurements + 8 monitoring measurements), and the simple mode can display four measurement
parameters. These displayed parameters can be arbitrarily selected from 25 power parameters according to the needs of users. In terms of
graphic display, a simple oscilloscope mode is provided to display waveforms for three parameters including voltage, current and power.
In addition, the measurement and analysis of each harmonic order of the measurement signal can be completely displayed by numerical
values or bar graphs. This power meter not only meets the needs of accuracy and legibility in process testing, but also meets the needs of
diverse measurement applications in R&D design and quality verification.
In addition, the performance of GPM-8310 in auxiliary measurement mechanism/function is also comprehensive. For the application of
measuring large voltage, the VT rate setting can be used with an external voltage Potential Transformer. For the measurement of large
current, the type of current transformer ~ voltage output type or current output type will determine the applied method. If it is a current
output type, it can be directly locked to the rear panel of the instrument and collocated with the CT rate setting to conduct measurement.
If it is a voltage output type, measurement can be conducted through the external current sensor input terminals (EXT1/EXT2) provided by
GPM-8310. Automatic level-changing can self-define the required level to save level-changing time. 10,000 lots of internal memories can
be used to store measurement data according to the update rate set by GPM-8310 or a user-defined time interval for subsequent analysis.
In terms of data retrieval and storage, GPM-8310 provides a variety of communication interfaces including RS-232C/ USB device (virtual
COM)/ LAN/ GPIB. Users can write programs to read the measurement results according to their habits or with existing system interfaces
and there is no need to procure interfaces. USB host supports GPM-8310 screen capture, internal record data access, and firmware
update. For the needs of external signal control or the use of data recorder to record data, GPM-8310 also provides an optional Digital I/O
(DA4) interface (must be installed before leaving the factory), which can be connected to an external controller such as PLC or a data
recorder to meet the application of automatic measurement or long recording.
Numerical (General)
Mode
Numerical (Simple)
Mode
Waveform
Mode
Harmonic (Bar Graph)
Measurement
Harmonic (Table Column)
Measurement
VARIOUS DISPLAY MODES
A.
GPM-8310 provides the numerical value display mode and the waveform
display mode, which help users to maximize the benefit of their
measurement. Under the numerical mode, there are the general mode and
the simple mode. The general mode has related measurement settings
and can simultaneously display 10 measurement parameters (2 main
measurements and 8 secondary measurements). The simple mode
displays only 4 measurement parameter results. The parameters in each
mode can be arranged and combined as required. Under the graphic
mode, a simple oscilloscope function is provided to display the waveforms
of three parameters including voltage, current and power. The horizontal
scale can be adjusted (from 25us/div ~ 1s/div according to the set data
update rate), and 3 magnification rates for waveform observation are also
provided for users to select. In the harmonic measurement, the
measurement results of each order of harmonics can be displayed by bar
graphs, and a specific observation order can be specified. The relevant
values of each order of harmonics (voltage/current/power/voltage
distortion ratio/current distortion ratio/power distortion ratio/voltage
phase angle/current phase angle) can be completely recorded and
displayed.