Data Sheet
GPT15000_Spec_E1.doc 2 OF 3
TIMER (Test Time)*
OFF, 0.3s~999.9s
TIMER Accuracy
(100ppm + 20ms)
WAIT TIME
0.0s~999.9s
GND
ON/OFF
DC WITHSTANDING
Output-Voltage Range
0.050kV~6.000kV
Output-Voltage Resolution
1V
Output-Voltage Accuracy
(1% of setting + 5V) [no load]
Maximum Rated Load
100W (5kV/20mA)
Maximum Rated Current
20mA (0.5kV< V ≦6kV)
2mA (0.05kV≦ V ≦0.5kV)
Voltage Regulation
(1% + 5V) [maximum rated load → no load]
Voltmeter Accuracy
(1% of reading + 5V)
Current Measurement Range
1μA~20.00mA
Current Best Resolution
0.1μA /1μA /10μA
Current Measurement Accuracy
(1.5% of reading + 3μA) when I Reading < 1mA
(1.5% of reading + 30μA) when I Reading ≧1mA
Current Offset
5μA Maximum
Window Comparator Method
Yes
ARC Detect
Yes
RAMP UP (Rise Time)
0.1s~999.9s
RAMP DOWN (Fall Time)
0.0s~999.9s
TIMER (Test Time)*
OFF, 0.3s~999.9s
TIMER Accuracy
(100ppm + 20ms)
WAIT TIME
0.0s~999.9s
GND
ON/OFF
INSULATION RESISTANCE
Output Voltage
50V~1200V dc
Output-Voltage Resolution
50V
Output-Voltage Accuracy
± (1% of setting + 5V) [no load]
Resistance Measurement
Test Voltage
Display Range
Measurement Range / Accuracy
50V≦V≦100V
0.1MΩ~ 10.00GΩ
0.1MΩ~1MΩ:±(5% of reading + 3 count)
1 MΩ~50MΩ:±(5% of reading + 1 count)
51MΩ~2GΩ:±(10% of reading + 1 count)
150V≦V≦450V
0.1MΩ~ 20.00GΩ
500V≦V≦1200V
0.1MΩ~ 50.00GΩ
0.1MΩ~1MΩ:±(5% of reading + 3 count)
1 MΩ~500MΩ:±(5% of reading + 1 count)
501MΩ~9.999GΩ:±(10% of reading + 1 count)
10GΩ~50GΩ:±(20% of reading + 1 count)*
Voltage Regulation
(1% + 5V) [maximum rated load → no load]
Voltmeter Accuracy
(1% of reading + 5V)
Short-Circuit Current
10mA max.
Output Impedance
2kΩ
Window Comparator Method
Yes
RAMP UP (Rise Time)
0.1s~999.9s
RAMP DOWN (Fall Time)
0.0s~999.9s
TIMER (Test Time)*
OFF, 0.3s~999.9s
TIMER Accuracy
(100ppm + 20ms)
WAIT TIME
0.0s~999.9s
GND
ON/OFF
Ground Bond
Output-Current
03.00A~32.00A ac
Output-Current Resolution
0.01A
Output-Current Accuracy
3A≦ I≦ 8A:±(1% of reading + 0.2A)
8A<I≦ 32A:±(1% of reading + 0.05A)
Test-Voltage
8Vac max (open circuit)