Specification
16
AC Withstanding Voltage
Output Voltage Range 0.050kV~ 5.000kV
Output Voltage Resolution 1V
Output Voltage Accuracy (1% of setting +5V) with no load
Maximum Rated Load 150VA (5kV/30mA)
Maximum Rated Current 30mA
0.001mA~10mA(0.05kV~0.5kV)
0.001mA~30mA(0.5kV~ 5kV)
Output Voltage Waveform Sine wave
Voltage Regulation ± (1% +5V)[Maximum rated load
Frequency 50 Hz / 60 Hz
Voltmeter Accuracy ± (1% of reading+ 5 V)
Current Measurement Range 0.001mA~30.00mA
Current Best Resolution 1uA (1uA~9.999mA)
10uA(10.00mA~30.00mA)
Current Measurement Accuracy ± (1.5% of reading+30uA)
ARC DETECT YES
RAMP TIME (Rise Time) 0.1~999.9S
FALL Time OFF~999.9S
WAIT Time OFF~999.9S
TIMER (Test Time) CONT², 0.3S~999.9S
GND ON/OFF