Electrical Safety Tester GPT-9600 Series QUICK START GUIDE GW INSTEK PART NO.
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied, reproduced or translated to another language without prior written consent of Good Will company. The information in this manual was correct at the time of printing. However, Good Will continues to improve products and reserves the right to change specification, equipment, and maintenance procedures at any time without notice. Good Will Instrument Co., Ltd. No.
Table of Contents Table of Contents SAFETY INSTRUCTIONS ................................................... 4 Safety Symbols ............................................................................................. 4 Safety Guidelines ......................................................................................... 5 Power cord for the United Kingdom ........................................................... 7 INTRODUCTION ...............................................................
GPT-9600 Series Quick Start Guide SAFETY INSTRUCTIONS This chapter contains important safety instructions that you must follow during operation and storage. Read the following before any operation to ensure your safety and to keep the instrument in the best possible condition. Safety Symbols These safety symbols may appear in this manual or on the instrument. WARNING Warning: Identifies conditions or practices that could result in injury or loss of life.
SAFETY INSTRUCTIONS Do not dispose electronic equipment as unsorted municipal waste. Please use a separate collection facility or contact the supplier from which this instrument was purchased. Safety Guidelines General Guideline CAUTION Do not place any heavy object on the instrument. Avoid severe impact or rough handling that leads to damaging the instrument. Do not discharge static electricity to the instrument. Use only mating connectors, not bare wires, for the terminals.
GPT-9600 Series Quick Start Guide Cleaning the GPT-9600 Operation Environment Disconnect the power cord before cleaning. Use a soft cloth dampened in a solution of mild detergent and water. Do not spray any liquid. Do not use chemicals containing harsh material such as benzene, toluene, xylene, and acetone.
SAFETY INSTRUCTIONS Power cord for the United Kingdom When using the safety tester in the United Kingdom, make sure the power cord meets the following safety instructions.
GPT-9600 Series Quick Start Guide INTRODUCTION This Quick Start Guide is intended as a fast introduction to operating the GPT-9600 Series Safety Testers. This Quick Start Guide assumes that the user is familiar with safety testers. For comprehensive instructions on the GPT-9600 Series, please see the User Manual, located on the accompanying CD. GPT-9600 Series Overview The GPT-9600 Series Safety Testers are AC/DC withstanding voltage and insulation resistance safety testers.
INTRODUCTION Main Features Performance Features Interface ACW: 5kVAC DCW: 6kVDC IR: 50V, 100V, 250V, 500V, 1000V Timer control Safety discharge Over temperature, voltage and current protection Pass, Fail, Test, High Voltage and Ready indicators PWM output (increased reliability) Interlock (configurable) Remote control start/stop interface terminal Signal I/O port for pass/fail/test monitoring and start/stop control/interlock Accessories Standard Accessories P
GPT-9600 Series Quick Start Guide Optional Accessories 10 Part number Description GHT-205 High Voltage Test Probe GHT-113 High Voltage Test Pistol GRA-417 Rack mount kit
INTRODUCTION Panel Overview Front Panel PASS/FAIL indicators Function keys Navigation keys Display AC / DC Withstanding Voltage / GPT-9603 Insulation Resistance Tester PASS FAIL ESC FIELD READY TEST CAUTION HIGH VOLTAGE START READY indicator HIGH VOLTAGE indicator HIGH VOLTAGE output terminal STOP RETURN REMOTE POWER POWER button EDIT SAVE TEST indicator 5.0 kVAC MAX. 6.0 kVDC MAX.
GPT-9600 Series Quick Start Guide OPERATION Status Modes This section describes the overall structure of the operating modes for the GPT-9600 Series safety testers. The testers have 6 status modes: (EDIT, READY, TEST, STOP, FAIL, PASS). The flow chart below describes how to navigate from mode to mode.
OPERATION READY Status Description When the tester is in READY status, it is ready to begin testing. Pressing the START button will begin testing and put the tester into TEST status. Pressing the EDIT key will put the tester into the EDIT status. Pressing the UTILIY key will enter the Common Utility settings. The READY status can also be used to set the testing mode. The testing mode can be set to run a single test at a time or to run two different tests sequentially. READY Status H = 2 2 .
GPT-9600 Series Quick Start Guide Go to EDIT Status Press the EDIT key. EDIT Set the Test Voltage When in the READY mode, turn the scroll wheel knob to set the test voltage. Start Testing Press the START button to begin testing. The tester will go into TEST status. WARNING WARNING START If Double Action is active, the START button must be pressed within 500ms after the STOP button was pressed to be able to start testing.
OPERATION High limit H = 2 2 . 0mA 0 00 A CW Low limit L = 0 . 0 0mA kV D CW EDIT Status T=174S mA I R ACW, DCW and IR test indicators Time setting ED I T MO D E Time soft-key Ground mode Choose Test Function Use the ACW, DCW or IR soft-keys to choose a test function. T I ME A CW Choose High/Low limits Use the FIELD key to bring the cursor to the H or L setting. Use the scroll wheel to edit the setting. Test Time Use the FIELD key to select the T setting.
GPT-9600 Series Quick Start Guide Exit EDIT status and Return to READY Status Press the ESC key. The tester does not save and returns back to READY status. ESC TEST Status Description TEST status is active when a test is running. Pressing STOP will cancel the test and put the tester into the STOP status. Waiting for the test to complete will result in a PASS or FAIL judgment. TEST Status H = 2 2 . 0mA 0 14 A CW kV D CW L = 0 .
OPERATION PASS/FAIL Result Results When the tester is allowed to run to completion, the test result is shown as a PASS or FAIL. When the PASS or FAIL results are shown on the screen, the PASS or FAIL indicators will also light up. Results FAIL Status H = 2 2 . 0mA 1 00 A CW kV D CW L = 0 . 0 0mA T=060S 23 0 mA FA I L I R MO D E Return to READY Status Press the STOP button once for a PASS judgment, or twice for a FAIL judgment.
GPT-9600 Series Quick Start Guide View FAIL Auto Mode Test Results For a Fail Judgment, press the STOP button once to turn off the buzzer tone. Then turn the scroll wheel left or right to view test results. Pressing the STOP button again will return the tester to READY status. STOP STOP Status Description STOP status is shown when a test does not finish running and has been stopped by the operator. When in STOP status, pressing STOP will return the tester to READY status. STOP Status H = 2 2 .
OPERATION Common Utility Settings Description The Common Utility menu is accessed by pressing the UTILITY key when the tester is in READY status. This utility controls the LCD and control settings. These settings are system wide.
GPT-9600 Series Quick Start Guide The tester will return to READY status. Cancel and Exit the Common Utility Menu To exit and cancel any changes, press the ESC key. ESC The tester will return to READY status. TEST Utility Settings Description The TEST Utility menu is accessed by pressing the UTILITY key when the tester is in EDIT status. The TEST Utility settings are configured for each test function (DCW or ACW) separately. The settings include: ARC MODE (ACW, DCW), and FREQUENCY (ACW).
OPERATION Cancel and Exit the TEST Utility Menu To exit and cancel any changes, press the ESC key. ESC The tester will return to EDIT status.