Manual

GPT-9000 Series Quick Start Guide
12
Add a Test to the AUTO Test (AUTO mode only)
Use the DOWN arrow key to bring the cursor to MANU
number.
Use the scroll wheel to select a MANU test number.
Press the ADD soft-key to add the test as the last step of
the AUTO test.
A D D
READY Status
Description
The tester is put into READY status when the
STOP button is pressed while in VIEW status.
When the tester is in READY status, it is ready
to begin testing. The READY indicator will be
lit when the tester is in READY status. Pressing
the START button will begin testing and put
the tester into TEST status. Pressing the
MANU/AUTO key will return the tester to
VIEW status.
For ground bond testing, the READY status is
also used to start the ZERO function which sets
the reference offset to compensate for the
resistance of the test leads.
Z E R 0 I R G B A CW D CW
E
F R E Q = 0 H z 6
0
3 01
A
G B V = V MT I E = 0 0 3 . 2 SR
. 0 m ER F =# 0 00
0 0 . 0 m IH ES = 3T
MM A UN N A_2M A N U = * * * - 0 0
m
READY status
AER
D
Y
Start Testing
Press the START button to begin testing.
The tester will go into TEST status.
START