Instructions

SAFETY INSTRUCTIONS
7
AC Input voltage range:
100 ~ 240 Vac
Frequency: 47 ~ 63 Hz
To avoid electrical shock connect the protective
grounding conductor of the AC power cord to
an earth ground.
The power switch that is included in the
instrument is not considered a disconnecting
device.
The power cord set is used as the disconnecting
device and shall remain readily operable.
Do not position the equipment so that it is
difficult to operate the disconnecting device.
Do not replace the power supply cord with an
improperly rated cord. A certified power supply
cord should not lighter than light PVC sheathed
flexible cord according to IEC 60227,
designation H05VV-F, and be rated for at least
3G 0.75 mm² (for rated current up to 10 A) or 3G
1.0mm² (for rated current over 10 A up to 16 A)
wire or larger, and the length of the cord that
does not exceed 2 m must be used.
Disconnect the power cord before cleaning.
Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals containing harsh material
such as benzene, toluene, xylene, and acetone.
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
Relative Humidity: 20%~ 80%, no condensation
Altitude: < 2000m
Temperature: 0°C to 40° C