Instructions

OPERATION
97
Advanced Settings
Using the Remote Sense Function from page 97
Preset Settings → from page 100
Using the Remote Sense Function
The ASR-2000 can be operated using local or remote voltage sense.
By default, the power supply is configured for local sense.
Ensure the output is off before handling the
remote sense connectors.
Use sense cables with a voltage rating exceeding
the isolation voltage of the power supply.
Never connect sensing cables when the output is
on. Electric shock or damage to the power supply
could result.
The remote sensing input
connector is located at the rear
panel of the ASR-2000.
Local Sense
When using local sense, the remote sensing input
terminal is not used. No compensation of any
possible voltage drop seen on the load cables is
performed. Local sense is only recommended
when the voltage drop is of no consequence. By
default, the power supply is configured for local
sense.
1. Check that the remote sense setting is disabled
(page 127).