User Manual User guide

APS-1102 LabVIEW Driver Manual
2-4
2.6 SOURce Subsystem
GW APS-1102 SOUR Curr Ave.vi
This sets the upper limit of the output average current.
<Input>
Average_Rn100V [A]
This sets the upper limit of the output average current in the 100 V range. The default
setting is “10.5”.
Average_Rn200V [A]
This sets the upper limit of the output average current in the 200 V range. The default
setting is “5.3”.
Rn_set
This specifies the range to be set. The default setting is “100 V (False)”.
<Output>
Output [A]
When Read = True, outputs the upper limit of output average current set for the APS-1102.
When Read = False, “99.9” is set.
GW APS-1102 SOUR Curr High.vi
This sets the upper limit of the output peak current (positive polarity).
<Input>
High_Rn100V [A]
This sets the upper limit of the output peak current (positive polarity) in the 100 V range.
The default setting is “42.0”.
High _Rn200V [A]
This sets the upper limit of the output peak current (positive polarity) in the 200 V range.
The default setting is “21.0”.
Rn_set
This specifies the range to be set. The default setting is “100 V (False)”.
<Output>
Output[A]
When Read = True, outputs the upper limit of output peak current (positive polarity) set for
the APS-1102. When Read = False, “99.9” is set.
GW APS-1102 SOUR Curr Low.vi
This sets the upper limit of the output peak current (negative polarity).
<Input>
Low_Rn100V [A]
This sets the upper limit of the output peak current (negative polarity) in the 100 V range.
The default setting is “–42.0”.