Compiling Arbitrary Waveform Input
Background
In order to generate arbitrary waveforms, it is
requested to use a specifically control software
on external PC which transfers data, via USB
interface, to the arbitrary waveform memory
with ASR-2000.
Note
Arbitrary waveforms cannot be changed when
output is on. To change arbitrary waveform,
make sure the output is off beforehand.
It is not allowed to compile the arbitrary
waveform memory directly from ASR-2000.
Only connecting with a PC with control
software via USB interface can complete it.
Memory
Arbitrary waveform memory count: 16
Arbitrary waveform length: 4096 words
Arbitrary waveform data: 16-bit binary (2’s
complement format)
Valid range of waveform data: -32767 to 32767
When a value greater than 32767 is input,
waveform data will be clipped to 32767. Also,
when a value less than -32767 is input, the
waveform data will be clipped to -32767.
Output Arbitrary
Waveform on
ASR-2000
1. Press Shift + Test to access the
Wave menu.
+
Alternatively, it is available to use
scroll wheel followed by the Enter
key to enter the Wave menu.