User Manual

TEST MODE FUNCTION
131
AC/DC Voltage
Range (ACV/DCV)
There are 2 voltage range settings:
HI 200V & LO 100V, which result in
varied ranges of ACV and DCV
values, respectively.
Fset (Frequency)
Sets the frequency of the step.
There are 3 secondary frequency
settings that determine how the
frequency is output.
CT: Sets the frequency level of the
step immediately to Fset values.
KP: Sets the frequency level to
“keep” the frequency of the
previous step.
SP: Linearly increases or decreases
the frequency from the end of the
previous step to the end of the
current step.
Note: it is available for AC+DC-INT
and AC-INT modes only.
Wave
Sets the outputting waveform of the
step. Up to 4 waves including sine,
square, triangle and arbitrary (1-16)
wave shapes are available.
Note: it is available for AC+DC-INT
and AC-INT modes only.
Jump To
The Jump To setting determines
which step to jump to at the end of
the step. If Jump To is turned off,
the unit will follow the Term (Step
termination) setting for the step.
Jump Cnt
Determines the number of times to
loop the jump step.