User Manual

ASR-2000 Series User Manual
38
Output Terminals
Background
The output terminals can be output from either
the front panel or from the rear panel. The
outputs are limited to 5 A / 2.5 A (ASR-2050),
10 A / 5 A (ASR-2100).
Supported Plugs
Multi-region terminal socket
Supported standards
IEC, North America, Japan.
EURO CEE type universal plug
WARNING
Dangerous voltages. Ensure that the power to the
instrument is disabled before handling the power
supply output terminals. Failing to do so may lead
to electric shock.
CAUTION
For the front panel output, the maximum output
voltage is 250 VAC and current is 10 A.
Front Panel
Output
Connection
1. The front panel has a multi-region power
socket depending on the socket type.
2. Insert the plug from the DUT into the socket.
(ASR-2000 or ASR-
2000R w/t GET
Series Box)
EURO CEE socket
IEC North America, Japan
WARNING
Dangerous voltages. Ensure output is off before
unplugging the plug from the front panel socket.