User Manual

ASR-2000 Series User Manual
150
Time
0.0001 ~ 999.9999s (Normal1,
Normal2 and Abnormal)
0.0000 ~ 999.9999s (Trans1 and
Trans2)
Note: For Trans1 and Trans2, it
supports a value of 0, which will
skip the step.
6. In order to adjust ACV voltage range between
HI and LO, it is required to set up outside of
the SIMULATE menu. Refer to the page 62 for
details. The selected range will be shown on
the top bar.
Range
Range
LO - 100V, HI - 200V
7. Go to the ACV setting and set the Vrms level
of the step. If you input an ACV value that is
not within the voltage range, the warning
message below will be shown.
Not applicable for Trans1, Trans2 and Normal2.
ACV
0.0 ~ 350.0V (Range 200V)
0.0 ~ 175.0V (Range 100V)