User Manual

ASR-2000 Series User Manual
188
Sensing Voltage Error
Remote sense connection wire is
abnormal or over maximum
compensation voltage. Press
"Shift + Cancel" to clear this
alarm.
Output Off
Startup Anomaly
Abnormal startup procedure.
Contact service center.
System Lock
External Sync
Frequency Error
The external synchronization
signal input frequency is out of the
allowance range. (40Hz ~
999.9Hz)
Output Off
SCPI Error
Communication with the SCPI
command error
Display Message
Only
Power ON Fail
Power ON Function Fail In Error
Mode or Range
IRMS Limit
The RMS current limiter is
activated. Press "Irms" to check
allowance set range
IPK Limit
The peak current limiter is
activated. Press " Shift + Irms" to
check allowance set range
Remote Sensing
Voltage Out of Range
The Sensing voltage limiter is
activated.
Display Message
Only
System Error (#)
System Error (1~15). Contact
service center.
Display Message
Only
Note
The system would be locked or output off automatically before the
error state is cleared.
Normal Messages
Description
Protection type
Setting Voltage
Limited
Setting voltage be limited, press
"Shift + V" to check allowance set
range
Display Message
Only
Setting Frequency
Limited
Setting frequency be limited, press
"Shift + F" to check allowance set
range
Display Message
Only
Keys Locked
All of keys are locked, except
output key. Long push Lock to
disable Keys Locked.
Display Message
Only
Keys Unlocked
All of keys are unlocked
Display Message
Only
Screen Saved
Screenshot be saved to USB
memory successful
Display Message
Only