User Manual

APPENDIX
197
Harmonic voltage
Effective value (rms)
Percent (%)
(AC-INT and 50/60
Hz only)
Range
Up to 40th order of the fundamental wave
Full Scale
175 V / 350 V, 100%
Resolution
0.1 V, 0.01%
Accuracy*8
Up to 20th
±(0.2 % of reading + 0.5 V / 1 V)
20th to 40th
±(0.3 % of reading + 0.5 V / 1 V)
Harmonic current
Effective value (rms)
Percent (%)
(AC-INT and 50/60
Hz only)
Range
Up to 40th order of the fundamental wave
Full Scale
5 A / 2.5 A, 100%
10 A / 5 A, 100%
Resolution
0.01 A, 0.01%
Accuracy*3
Up to 20th
±(1 % of reading +
0.1 A / 0.05 A)
20th to 40th
±(1.5 % of reading +
0.1 A / 0.05 A)
Up to 20th
±(1 % of reading +
0.2 A / 0.1 A)
20th to 40th
±(1.5 % of reading +
0.2 A / 0.1 A)
*1
The voltage display is set to RMS in AC/AC+DC mode and AVG in DC mode
*2
AC mode: For an output voltage of 17.5 V to 175 V / 35 V to 350 V and 23 °C ± 5 °C.
DC mode: For an output voltage of 25 V to 250 V / 50 V to 500 V and 23 °C ± 5.
*3
An output current in the range of 5 % to 100 % of the maximum current, and 23 °C ± 5 °C.
*4
An output current in the range of 5 % to 100 % of the maximum peak current in AC
mode, an output current in the range of 5 % to 100 % of the maximum instantaneous
current in DC mode, and 23 °C ± 5 °C. The accuracy of the peak value is for a
waveform of DC or sine wave
*5
For an output voltage of 50 V or greater, an output current in the range of 10 % to 100 %
of the maximum current, DC or an output frequency of 45 Hz to 65 Hz, and 23 °C ± 5 °C.
*6
The apparent and reactive powers are not displayed in the DC mode.
*7
The reactive power is for the load with the power factor 0.5 or lower.
*8
An output voltage in the range of 17.5 V to 175 V / 35 V to 350 V and 23 °C ± 5 °C.