User Manual

ASR-2000 Series User Manual
74
Setting the Output Frequency & Signal
Background
The FREQ and SIN settings set the frequency of
the output. Before setting the frequency, set the
frequency limit.
Steps
1. Press the F key to access the FREQ
or SIG parameter depending on
varied modes.
Also, it is available to use the
scroll wheel followed by the Enter
key to make the FREQ or SIG
parameter selectable as well.
2. Set the frequency or signal with the scroll
wheel or with the F1 ~ F4 soft-keys.
AC+DC-INT, AC+DC-ADD
FREQ
Range
1.00 ~ 999.9 Hz
Soft-keys
DEF1, DEF2, MAX, MIN
AC-INT, AC-ADD
FREQ
Range
40.00 ~ 999.9 Hz
Soft-keys
DEF1, DEF2, MAX, MIN
AC+DC-Sync, AC-Sync
SIG
Option
LINE, EXT
3. Press Enter key to confirm the frequency or
signal setting.
Defined Settings
The DEF1 and DEF2 settings are user defined
settings. By default they are set to 50.00 Hz and
60.00 Hz, respectively. The MAX and MIN soft-
keys set the frequency to the maximum and
minimum, respectively.