Example of I
Setting in the
DC-INT
I setting
Soft-keys setting
F3
F4
F1
IRMS & I Limit
On/Off
Almost identical with the concept of previous IPK Limit
function, the IRMS/I Limit function keeps the IRMS/I
value within the certain limit when the predefined value
is reached. However, due to RMS calculation, the unit
requires approximate 200ms of detect time before
starting the adjustment process so that the IRMS/I limit
can be well maintained. If, on the other hand, this
function is turned off, the output will be disabled
instantly when IRMS/I Limit off level is reached.
IRMS & I Limit
On
3. After entering the either IRMS or I
setting, press F1 soft key to turn
IRMS Limit function On.
IRMS & I
Limit ON:
T
Kept within IRMS & I
Limit On Level
I
Detect Time
Approx. 200ms
Adjustment
process
Load On