COMMAND SET
43
:TEMPlate:MAX
Set
Query
Description
Sets or returns Go-No Go Max template.
Same as: Utility key → F5 (More) → F3 (Go-NoGo
Menu) → F1 (Template Edit) → F1 (Template Max) →
F2 (Source)
Syntax
< Long >
< Short >
:template:max <NR1>
:template:max?
:temp:max <NR1>
:temp:max?
Parameter
<NR1>
Waveform used for Go-NoGo maximum
template
0
Reference waveform A
1 ~ 20
Internally stored waveform W1 ~ W20
Before this
command...
Go-NoGo normal template mode has to be selected
prior to this command.
:TEMPlate:MODe 0 (for details, see pagexx.)
Example
:template:mode 0
:template:max 1
Select normal template
mode, then select
internal waveform W1
as Max template