Manual
Table Of Contents
![](/manual/gw-instek/gbs-1000-series-programming-manual/manual-english/images/img-80.png)
COMMAND SET
81
1
>
2
=
3
≠
Example
:trigger:pulse:mode 1
Sets the triggering
condition to “>”.
:TRIGger:PULSe:TIMe
Set
Query
Description
Select the time value for pulse width. The setting
range is from 20ns to 10s.
Same as: Trigger MENU key → F4(When) →
Variable knob. (When trigger mode = pulse)
Syntax
< Long >
< Short >
:trigger:pulse:time<NR3>
:trig:pulse:time?
:trig:puls:tim<NR3>
:trig:puls:tim?
Parameter/Return
parameter
<NR3>
Pulse width in seconds
Example
:trigger:pulse:time?
>1.000E+01
The pulse width is 10
seconds.
:TRIGger:REJect
Set
Query
Description
Sets or queries the frequency rejection mode.
Same as: Trigger MENU key →
F5(Slope/Coupling)→F3(Rejection).
Syntax
< Long >
< Short >
:trigger:reject < 0 | 1 | 2 >
:trig:reject?
:trig:rej <0 | 1 | 2 >
:trig:rej?
Parameter/Return
parameter
<NR1>
0
Off
1
Low frequency rejection